The auxiliary light may be used for general illumination of the surgical area or alternatively for
illumination of the instrument tray. To adjust the position of the auxiliary light, loosen the clamp
knob which is located underneath the universal joint. Hold the auxiliary light with one hand and
manoeuvre the light until it is pointing in the desired direction. Tighten the clamp knob firmly.
To tilt the head assembly up or down, simply rotate the tilt knob anti-clockwise or clockwise
accordingly. Note that the entire head assembly will tilt, not just the eyepieces. To ease this
operation, support the weight of the microscope head with one hand while using the other hand to
rotate the knob.
#
To zoom the image in or out, depress the appropriate foot control pedal. The total zoom range is
between 4x and 25x magnification. Sterilisable covers are provided for fitting over the manual zoom
knob when sterile use is required.
To focus the microscope up or down, the appropriate foot control pedal. The total focus range is
50mm. For optimum microscope use, leave the microscope head in such a position to allow
approximately 25mm of focus range in each direction. Sterilisable covers are provided for fitting over
the manual focus knob when sterile use is required.
If the manual controls are to be used (for example where no power is available) first disconnect the
foot control. This will significantly reduce wear on the focus and zoom motors, and enable easier
turning of the manual focus and zoom knobs.