USE IN LEGAL FOR TRADE APPLICATIONS : 

 

4.1 Introduction 

e

Nod3-

is a module making a part of an instrument. It is intended to be integrated : 

-

 

In a non automatic weighing instrument (

NAWI

) or 

-

 

In an automatic weighing instrument (

AWI

) of category catchweigher. 

 

4.2  Legal for trade parameters 

See relative sections in the following documents : 

-

 

ModBus RTU communication

 Ref. 165704 

  § 2.11 Metrological version number 

  § 2.12 Legal for trade 

  § 2.37 Status register 

-

 

SCMBus communication

 ref. 165706 

  § 4.3 Legal for trade settings 

  § 1.4 Status bytes 

-

 

CANOpen® communication protocol

 Ref. 165708 

  § 3.2.16 Legal for trade switch 

  § 3.2.17 Legal for trade indicators 

  § 3.2.32 Current measurement 

  § 3.2.33 Current measurement status 

 

4.3 Sealing 

The module 

e

Nod3-

C

 has a physical sealing device and a software sealing device. 

4.3.1 

Physical sealing : 

The physical sealing is the one of  the module box. It comprises two self-adhesive stickers self-
destructive when removed or a device with sealing screws with lead and sealing wire. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

4.3.2 

Software sealing : 

On the whole weighing instrument the value of the event counter as well as the CRC value may be 
displayed. 
The whole weighing instrument has a marking area where the values of the event counter and of 
the CRC recorded after the last official verification.These marked values shall be identical to these 
displayed on the terminal. When these values do not match, this part of the sealing device is 
considered as broken. 

4.4  Specific requirements : 

The legal metrological software version may be displayed on the terminal. 
There are also some characteristics in the status related to measurements that need to be 
displayed on the terminal too. 

 

Sealing with self-adhesive stickers self-

destructive when removed put on 2 

opposite sides of the box, with a part on 

the top and a part on the side of the box. 

Sealing with lead and sealing wire, 2 

opposite screws have to be sealed. 

Содержание eNod3-C

Страница 1: ...eNod3 C Digital Transmitter eNod3 C User s instructions NU eNod3C E 0911_165702 I 1 22 User s instructions...

Страница 2: ...on E 04 08 new inputs assignement allow new cycle and peak detection cycle decription modified F 11 08 document title modified eNod3 eNod3 C band stop filter and automatic zero correction in checkweig...

Страница 3: ...s attached to an operating mode 13 7 OUTPUTS FUNCTIONING 14 7 1 Assignment of outputs 14 7 2 General functions 14 7 3 Functions attached to an operating mode 14 8 SET POINTS 15 9 FILTERS 16 10 TRANSMI...

Страница 4: ...tocols Each module is provided with 2 logical inputs and 2 logical outputs authorizing synchronization of functions with automation and alarm management SCAIME provides the eNodView software to facili...

Страница 5: ...evel 10mA 24V mA insulation voltage 2500 Vrms Logical outputs number 2 type opto insulated static relays max current 40 C 0 4 A max voltage in open state 55 V resistance in ON state 2 insulation volta...

Страница 6: ...rogrammable functions acquisition of zero taring zero tracking physical or theoretical calibration slope correction non linearity polynomial correction low pass band stop and self adaptive digital fil...

Страница 7: ...onnected 2 2 Connection to load cell s eNod3 C supplies power to the load cells 5 VDC Up to four 350 load cells can be connected in parallel eNod3 C allows the use of 4 or 6 wire load cells 4 wire loa...

Страница 8: ...PB 2 3 2 Digital outputs Output characteristics opto insulated static relays Max current 40 C 0 4A Max voltage in the open state 55V Resistance in the ON state 2 Insulation voltage 2500 Vrms A light...

Страница 9: ...onding jumper OFF For an RS422 or RS485 full duplex communication use the four TB TA RB and RA pins The corresponding jumper must be in place ON which is the default case on delivery The RS232 interfa...

Страница 10: ...management on the line This operating mode is only available in transmitter functioning mode The measurements transmission frequency depends on the serial baud rate thus transmitting 100 meas s is imp...

Страница 11: ...al sealing The physical sealing is the one of the module box It comprises two self adhesive stickers self destructive when removed or a device with sealing screws with lead and sealing wire 4 3 2 Soft...

Страница 12: ...e of calibration can be done with 1 2 or 3 known references Theoretical adjustment by setting the load cell sensitivity and capacity Correction of the initial calibration value 5 2 Linearization corre...

Страница 13: ...er or both inputs may be assigned to the tare function The tare is affected by a stability criterion that can be parameter defined The tare will be triggered on a rising or a falling edge depending on...

Страница 14: ...ing measurement stability defective measurement outputs can be assigned to copying measurement faults These faults are also coded in the status word attached to the measurements There are 4 of them ou...

Страница 15: ...alues of these set points may be assigned either to gross measurement regardless of the functionning mode net measurement regardless of the functionning mode maximum value in peak control mode minimum...

Страница 16: ...erworth type low pass filter a low pass digital filter can be applied as an output of the A D converter The filter order is configurable available values are 2 3 or 4 and the coefficients that define...

Страница 17: ...ampling period setting The request can apply to gross measurement net measurement measurement in A D converter points Note This is very similar to operation of the Measurement window through an input...

Страница 18: ...only input 2 is operational if both inputs are assigned to the start checkweigher cycle function by a trigger level Fig 9 when the load cell signal reaches the specified value Then during a stabiliza...

Страница 19: ...SCMBus protocol the output state changes when it is read 11 2 2 With the SCMBus protocol In Checkweigher automatic transmission mode when the cycle is finished the result is automatically sent through...

Страница 20: ...d within a configurable interval around the zero calibration and during a configurable period Some other criteria are also taken in account they are Number of measurements complying with correction ra...

Страница 21: ...the net current value and the Peak to peak value to zero Fig 11 12 2 Triggered operating mode A cycle is managed It can be managed through an input assigned to a measurement window Fig 13 through a t...

Страница 22: ...Set points can be assigned with either Max or Min or Peak to peak values or to the gross or net measurement 12 4 Other possible outputs assignments cycle in progress defective measurements Fig 12 Fig...

Отзывы: