sauter TVM 10kN120N Скачать руководство пользователя страница 7

 

TVM-N-BA-e-2020 

 

Application 

8.1 

Check before starting the measurement / test 

  Wiring, switching on Display flashes 5 times 

  Test the movement without the test piece, manually actuating the limit 

switches to test their function. 

  Test of the automatic movement. Press the Manual/Auto Mode button, Auto 

Mode indicator lights up. Set cycles (avoid setting "1"), start test run with Up or 
Down button. At the end of the cycles, the test bench stops and emits an 
alarm tone 3 times, test finished. 

8.2 

Speed setting 

The speed can be adjusted continuously up to the maximum. The set speed can be 
read off the display. 

8.3 

Presettable cycles 

A number of cycles can be preset on the test bench. The preset value is displayed in 

the lower area. It can be set

 with the keys Counter 

, Counter 

and Select 

counter position. The "run" number is displayed in the upper area. The counter can 

be

 reset with the Zero key. 

8.4 

RS 232 connection 

The test stand has two 9-pin connectors to connect a force gauge and one connector 
for communication with the PC. The test stand can be operated with SAUTER AFH 
software. This allows the motion control and number of cycles to be set directly on 
the PC. The software can be used to evaluate the data in terms of force-time or 
force-displacement. The test stand can be controlled at the connection for the force 
measuring device using an FH series force measuring device to prevent overload 
(using the Stop parameter on the FH series force measuring device). 

8.5 

Limit switch 

In manual mode, movement stops when the limit switches are reached. In Automatic 
mode, movement stops at the Perimeter Switch for about 5 seconds and then 
continues in the opposite direction. In order to ensure that the test/examinations run 
smoothly, you should ensure that you align the boundary rings very precisely so that 
the test object or test equipment is not destroyed if the path is too long/short. 

 

 

Содержание TVM 10kN120N

Страница 1: ...72336 Balingen e mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction manual motorized test bench SAUTER TVM N Version 2 0 01 2020 GB TVM N BA e 202...

Страница 2: ...Scope of delivery 3 3 Weight and dimensions 3 4 Check before use 3 5 Possible applications 4 6 Technical data 5 7 Control panel 6 8 Application 7 8 1 Check before starting the measurement test 7 8 2 S...

Страница 3: ...SAUTER TVM Power cord Operating instructions Accessories depending on model 3 Weight and dimensions Test bench TVM 5000N230N TVM 5000N230NL TVM 10KN120N TVM 20KN120N TVM 30KN70N Dimension LxWxH 410x26...

Страница 4: ...the setting options of the FH force gauge can be used for example to protect the test stand from overload with the STOP value Choose the right test stand with regard to the maximum force you require A...

Страница 5: ...00 N 30 000N Speed range 10 230 mm min 10 230 mm min 30 120 mm min 30 120 mm min 5 70 mm min Speed accuracy 1 100 mm min 2 mm min 100 mm min 10 Maximum travel distance 200mm Maximum number of cycles 1...

Страница 6: ...ol knob Regulation of the lifting speed Manual Auto Mode Choice between manual or automatic movement Display of preset cycles With the help of the counters counters and selection of counter position a...

Страница 7: ...The counter can be reset with the Zero key 8 4 RS 232 connection The test stand has two 9 pin connectors to connect a force gauge and one connector for communication with the PC The test stand can be...

Страница 8: ...trained and experienced personnel In particular it should be avoided that forces are applied to the measuring instrument by the test bench which exceed the maximum load Max of the test bench or measur...

Страница 9: ...ere as an example with a FH 500 10 2 with external load cell TVM 5000N230N NL and TVM 10KN120N Mounting bracket AFM 41 with 4x M6x35 black with washer and spring washer on crossbar Screws The external...

Страница 10: ...AFM 41 mounting bracket with an M12x80 screw for TVM 20KN and TVM 30KN included in the scope of delivery Here as an example with a FH 20K 10 4 Wiring of the test bench model independent The force gau...

Страница 11: ...tension cable to PC 10 5 Cabling TVM with a force measuring and length measuring device Wire the force gauge wiring as described in previous points of section 10 Connect the linear encoder to the PC v...

Страница 12: ...12 TVM N BA e 2020 11 Technical drawings...

Страница 13: ...H mm I mm TVM 5000N230N 250x380x410 635 32 220 185 400 165 TVM 5000N230NL 250x380x410 1120 32 220 190 400 165 TVM 10kN120N 250x380x410 1120 32 220 180 400 165 TVM 20kN120N 295x545x480 1135 35 295 200...

Отзывы: