          TF_TG-BA-e-1813 

4  Measurement procedure 

 
4.1 The power key 3-3 has to be pressed to switch on the instrument. `0` appears on 
the display 3-2. 
Both instruments, TF and TG, will recognize the mode of the last measurement itself 
by the symbol appearing in the display `Fe` (= F) for ferrous metals or `NFe` (= N) for 
non-ferrous metals. The instrument changes into the automatic measurement mode, 
which also is able to recognize and assign also the zeroing plate or any other base 
material.      
       
4.2 The sensor 3-1 has to be placed onto a coating layer to be measured. The read-
ing on the display is the thickness of the coating layer. This can be corrected by 
pressing the Plus- key 3-4 or the Minus- key 3-5. For doing this, the sensor should be 
kept away from the measured object or the base plate for at least 5 cm.           
                      
4.3 To perform the next measurement, the sensor 3-1 has to be lifted for more than 
1cm, “0” will be displayed and step b) has to be repeated. The instrument memorizes 
the continuous measured value automatically with statistic measurement times. 
Meanwhile, the Max, Min and average value will be displayed.           
    
4.4 To avoid any appearing incorrect measurement values, we recommend adjusting  
the instrument before starting with the measurements, as described in chapter 9. 
 
4.5 To change the measurement unit from „µm“ to “mil” or vice verse, the Power-key 
3-3  has to be pressed and not released until “UNIT” is shown on the display.  Then 
the same key 3-3 (this time Zero-key) has to be pressed. 
                
4.6 To change the measuring mode from `single` to `continuous`  or vice verse, the 
Power-key / Zero –key 3-3 has to be pressed and not released until `SC` appears on 
the display. Then the Zero- key 3-3 (the same one) has to be pressed.  
The symbol “STATS” represents the continuous mode and `S` represents the single 
mode.               

5  Statistics 

 
This instrument calculates and displays a statistical analysis of readings while the 
measurements are taken. 
The statistics available are: 

  Last value 

  Average value, marked by AVE 

  Highest reading marked by Max 

  Lowest reading marked by Min 

  Number of measurements taken 

Содержание TF 1250-0.1FN

Страница 1: ...info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TF TG Version 1 3 01 2018 GB Model TF 1250 0 1FN Model TG 1250 0 1FN TF_TG BA e 1813 PROFESSIONAL MEASURING ...

Страница 2: ...ons do not hesitate to call our service number Models available TF 1250 0 1 FN TG 1250 0 1 FN Summarize 1 Features 3 2 Specifications 4 3 Front panel description 5 4 Measurement procedure 6 5 Statistics 6 6 How to store and recall readings 7 7 How to delete readings 7 8 Transfer of readings to a PC 7 9 Calibration 8 10 Battery replacement 8 11 Adjustment foils 8 12 Correct handling at Coating Thic...

Страница 3: ...uer layers etc These layers should be located on magnetic metals e g steel iron nickel etc It is often used to measure the thickness of galvanizing layer lacquer layer porcelain enamel layer phosphate layer copper tile aluminium tile some alloy tile paper etc In N mode the coating thickness of non magnetic coatings is measured which are located on non magnetic metals It is used to measure anodizin...

Страница 4: ...t PC connection by RS 232C interface Power supply 2 x 1 5V AAA UM 4 batteries Operating conditions Temperatures 0 up to 50 C Humidity 95 Dimensions 126 x 65 x 35 mm 5 0 x 2 6 x 1 6 Inch Weight ca 81g without batteries Scope of delivery Carrying case Operation manual TF 1250 0 1FN Inbuilt sensor F as well as inbuilt sensor N base material is recognized automatically TG 1250 0 1FN External sensor FN...

Страница 5: ...ng key READ key 3 7 Delete key DEL key 3 8 Socket for RS 232 interface 3 9 Battery cover 3 10 Wrist ring 3 11 Battery indicator 3 12 Symbol for ongoing measurement 3 13 Last reading browsing value 3 14 Indication for single measurement mode 3 15 Max value indicator 3 16 Min value indicator 3 17 Browsing state 3 18 Measuring state 3 19 Unit 3 20 Average indicator 3 21 Base material indicator 3 22 C...

Страница 6: ...epeated The instrument memorizes the continuous measured value automatically with statistic measurement times Meanwhile the Max Min and average value will be displayed 4 4 To avoid any appearing incorrect measurement values we recommend adjusting the instrument before starting with the measurements as described in chapter 9 4 5 To change the measurement unit from µm to mil or vice verse the Power ...

Страница 7: ...s on the display it means that there is no reading to delete any more 6 4 The Zero key 3 3 has to be pressed to quit the measurement state 7 How to delete readings 7 1 To delete a reading on the display the DEL key has to be pressed no matter whether you are located in the measurement state marked SV or in browsing state marked RD Browsing state can be entered by pressing the READ key and meas ure...

Страница 8: ... distance from the border of the base plate 9 4 The sensor 3 1 has to be pressed carefully onto the calibration foil and then lifted The reading on the display is the value measured This can be corrected by pressing the Plus key 3 4 or the Minus key 3 5 while the sensor is removed from the base plate or the measured object 9 5 Step 9 4 has to be repeated until the desired accuracy is achieved 10 B...

Страница 9: ... beginning 13 2 Sensors will eventually wear off Life of the sensor will depend on the number of measurements taken and how abrasive the coating is Replacement of a sensor should only be performed by qualified staff 14 Restore factory settings 14 1 In the following cases it is recommended to restore factory settings The instrument does not measure any more Measurement accuracy is degraded caused b...

Страница 10: ...the accuracy This value should only be adjusted by professional persons Generally said The bigger the value of Ln the smaller the reading on the display for the same coat ing thickness Only a small change on the value of Ln causes a big change in the reading of the upper measurement range at 500µm 20 mil The value of Ln has to be adjusted as follows Press the Power key It lasts about 14 seconds fr...

Отзывы: