4.    Measuring procedure

 

 

4.1 The power- key is to be pressed to turn on the power.         
      `0` appears on the display 3-2. 
      

Remark: 

The instrument will auto- calibrate, as soon  

      as it is switched on. 
      Please pay attention to the sensor. During the initial  
      calibration it should be neither in direct surroundings 

      

of the base plate nor other magnetic materials. 

 
4.2 The sensor is to be placed on the coating to be  
      measured.

 

      The coating thickness is now shown on the display. 
 
4.3 To perform the next measurement the sensor has to  
      be lifted for more  than 1cm off the base material 
      and step 4.2 is to be repeated. 
 
4.4 In case of inaccuracies to the measurement result, it   
      is recommended to calibrate the instrument before 
      measuring as described in 6. 
 
4.5 The instrument can be switched off by the 
      Power- on/ Power- off key 3-6. The gauge 
      switches off automatically 2 minutes after the  
      last operation. 

 
5. OFFSET-
Accur 
 

With this instrument the possibility is given to improve 
the measurement result essentially by using the 

OFFSET- 

Accur function. Therefore it is necessary to calibrate the 
gauge with a reference coating in the typically measured   
range. This adjustment can also be done with the 
calibration foils, included in the delivery. Ideally this adjust- 
ment should be done on the base material which is used 
for the  real measurement instead on the base plate  
included in the delivery. 
 
5.1 The result of the initially measurement is still shown on        

      

the display (as performed in 4.2). 

 
5.2 The sensor is to be removed from the test object. Then  
      the reading on the display shall be corrected by 
      pressing the Plus- key or the Minus- key.  

 
6. Calibration

      

 
6.1 Zero calibration: The sensor has to be placed on the   
      base plate or another uncoated nonmagnetic material. 
      On the display `0` is shown when pressing the 
      Zero- key. 
      The Zero- key may not be pressed if the sensor is not 
      placed on the base plate or on any other uncoated  
      base material. 
 
6.2 An applicable calibration foil is to be chosen adequate  
      to the typical measurement range. 
 
6.3 The chosen calibration foil has to be placed onto the     
      base plate or another uncoated base material. 
 
6.4 The sensor 3-1 has to be placed carefully onto the  
      calibration foil and then it has to be lifted.        
      The result now appears on the display. 
      It shall be corrected by pressing the Plus- key 3-4 or   
      the Minus- key 3-5. For doing this, the sensor must be  
      removed from the base plate or the material to be 
      measured.   
    
6.5 Step 6.4 is to be repeated until the measurement       
      accuracy has been realised. 
 

7. Battery replacement

 

 
7.1 The batteries are to be replaced if the battery 
      voltage is less than 4.8V. In this case the battery  
      symbol ``+-`` appears on the display. 
 
7.2 The battery cover is to be removed (Fig.1, 3-6) from  
      the instrument and the batteries are to be taken off. 
 
7.3 The batteries (4×1.5V AA/UM-3) are to be installed    
      correctly into the case. 
 
7.4 If the instrument is not to be used for an extended  
      period, batteries are to be extracted. 

                      
8. Calibration foils 
 

As accessory the instrument includes a calibration foil  
set and the base plate.  
 

9. Correct handling of coating thickness measurement      
    with external sensors 

 
The sensor has to be touched at the lower pole segment 
and it has to be pressed slightly onto the test object. 