And the other way round, LN value should be decreased when the readout value in
the lower range is correct (e.g. 51 µm), and in the upper it is too low (e.g. 432 µm).
C. The processes in A and B should be repeated until the readout for every adjust-
ment film is satisfactory in terms of its accuracy.