background image

 

Sauter GmbH

 

Ziegelei 1 
D-72336 Balingen 
E-Mail: [email protected]

 

Tel:  +49-[0]7433- 9933-199 
Fax: +49-[0]7433-9933-149 
Internet: www.sauter.eu 

 

 

Instruction Manual 

HMP 

 

HMP-BA-e-1211 

 

7

 

 

5.10   Auto Power Off 

The  gauge  features  an  Auto  Power  Off  function  designed 
to  conserve  battery’s  life.  If  the  hardness  tester  is  idle 
(neither measuring nor any key operation) for 5 minutes, it 
will  turn  itself  off.  Before,  the  LCD  display  will  continue 

flashing for 20 seconds. Exept the 

 

key, any key can be 

pressed to stop the twinkle of the LCD display.  
Like this, the operation of powering off at the moment can 
be interrupted. 
While  the  voltage  of  the  battery  is  too  low,  <E00>  will  be 
displayed. Then the gauge is automatically powered off. 
 

5.11 Battery Replacement 

Two  AA  size  Alkaline  batteries  are  necessary  as  power 
resource.  After  several  hours’  usage,  the  battery  symbol 

on the screen will be shown as  

 . The less dark parts 

are  indicated,  the  more  close  to  be  filled.  When  the 
capacity  runs  out,  the  battery  symbol  will  be  shown  as 

    and  the  symbol  will  begin  to  flash.  Now  batteries 

have to be replaced. 

Pay attention to the polarity of the batteries! 
The  batteries  have  to  be  taken  out  if  the  instrument 
hasn’t been used for a longer period of time. 
 

5.12   Connection to PC 

The hardness tester HMO is equipped with a RS-232 serial  
port.  Using  the  accessory  cable  (the  cable  and  following 
referred  software  are  optional  parts)  the  gauge  has  the 
ability  to  get  connected  to  a computer  or  external  storage 
device. Measurement data stored in memory of the gauge 
can be transferred to PC with the RS-232 port. 
Detailed information of the communication software and its 
usage, please refer to the software manual. 
 

5.13   Error Code Reference 

 

 
6.       Maintenance & Service 
6.1     Impact Device Maintenance 
 

 

After  1000  to  2000  times  of  usage,  the  impact 
body and the guide tube have to be cleaned with 
the  provided  nylon  brush.  When  cleaning  the 
guide  tube,  the  support  ring  has  to  be 
unscrewed first. Then the impact body has to be 
taken out. The nylon brush has to be spiralled in 
counter-clock  direction  into  the  bottom  of  the 
guide  tube.  It  has  to  be  taken  out  and  repeated 
for  5  times.  At  last,  the  impact  body  and  the 
support ring have to be installed again. 

 

The impact body has to be released after use.

 

 

Any  lubricant  is  absolutely  prohibited  inside  the 
impact device.

 

 
6.2     Instrument Maintenance Program 

*  If  the  standard  Rockwell  hardness  block  is  used  to  test 
and if all the error is bigger than 2 HRC, the ball top of the 
impact  device  or  the  impact  object  may  have  to  be 
changed. The invalidation may be caused by abrasion. 
*  If  any  other  abnormal  phenomena  appear  to  the 
hardness  tester,  nothing  may  be  dismantled  or  any  fixed 
parts may not be adjusted by the user himself.  
The warranty card has to be filled in and sent to SAUTER 
GmbH. The warranty service will be performed within a few 
working days. 
 

6.3     Fault Analysis & Clearance 

 

 

6.4     Note of Transport and Storage Conditions 

*  The  instrument  has  to  be  kept  away  from  vibration, 
strong  magnetic  field,  corrosive  medium,  dampness  and 
dust. Storage in ordinary temperature. 
 

DESIGNED IN REGARD TO THESE STANDARDS: 

 

ASTM A956 

 

DIN 50156 

 
10. Declaration of conformity 

 

  

Содержание HMP

Страница 1: ...Program 6 3 Fault Analysis Clearance 6 4 Note of Transport and Storage Conditions APPENDIX Table 1 Table 2 Table 3 Table 4 1 Overview 1 1 Advantages Wide measuring range Based on the principle of LEE...

Страница 2: ...y Remarks Stand ard Config ura tion 1 Main unit 1 2 D type impact device 1 With cable 3 Standard test block 1 4 Cleaning brush I 1 5 Small support ring 1 6 Alkaline battery 2 AA size 7 Manual 1 8 Inst...

Страница 3: ...To change the impact times in one group the key has to be pressed The impact times item will start flashing when the key is pressed Then the impact times value can be adjusted as desired by the keys a...

Страница 4: ...the hardened layer should also conform to Table 3 Coupling Light weight samples must be firmly coupled with a heavy base plate Both coupled surfaces have to be flat and smooth there may not rest any...

Страница 5: ...he type of impact device will be automatically detected while powering on This information will be displayed on the screen Users should take notice to this type After a few seconds the screen will exi...

Страница 6: ...me during data logging 5 6 2 Clearing selected File Group It may be required to delete a file from the instrument s memory This procedure is outlined in following steps 1 The key has to be pressed to...

Страница 7: ...Measurement data stored in memory of the gauge can be transferred to PC with the RS 232 port Detailed information of the communication software and its usage please refer to the software manual 5 13...

Страница 8: ...tool steel HRC 20 4 67 1 19 8 68 2 20 7 68 2 22 6 70 2 HV 80 898 80 935 100 941 82 1009 Stainless steel HRB 46 5 101 7 HB 85 655 HV 85 802 Grey cast iron HRC HB 93 334 92 326 HV Nodular cast iron HRC...

Страница 9: ...e surface Ra 1 6 m 1 6 m 0 4 m 6 3 m 1 6 m Min weight of sample Measure directly Need support firmly Need coupling tightly 5kg 2 5kg 0 05 2kg 5kg 2 5kg 0 05 2kg 1 5kg 0 5 1 5kg 0 02 0 5kg 15kg 5 15kg...

Страница 10: ...sting cylindrical outside R25 R50 4 HZ11 13 For testing cylindrical inside R11 R13 5 HZ12 5 17 For testing cylindrical inside R12 5 R17 6 HZ16 5 30 For testing cylindrical inside R16 5 R30 7 K10 15 Fo...

Отзывы: