Use the keys
【 】
und
【 】
wird to move the cursor to
【
About Software
】
bewegt.
Press the
【
ENTER
】
wird key to go to
【
About Software
】
zu.
Here the user receives information about the hardness tester and its software on the
screen.
However, the software version and the built-in software identification can change
continuously, as they are subject to the continuous improvement process (upgrading)
of the device.
7.11 Calibration / Adjustment
The hardness tester as well as the rebound sensor must be calibrated using the Leeb
hardness calibration block before first use. This is included with the HK-DB, but not
with the HK-D. It can also be purchased as an option. The hardness tester must also
be calibrated if the device has not been used for a long period of time. It is sufficient to
calibrate each individual rebound sensor once in conjunction with the display unit;
recalibration is no longer necessary after a rebound sensor has been replaced.
The key
【
】
und is pressed simultaneously with the key
【
ENTER
】
to enter the
system. The Calibration software screen appears:
The position of the rebound sensor should be
【
】
sein.
5 points are measured in vertical direction on the Leeb hardness calibration block.
The average value is shown after these measurements.
Use the keys
【 】
und
【 】
wird to enter the nominal value.
Confirm this step with the key
【
ENTER
】
wird.
Press the
【
ESC button to cancel
】
kann. Adjustment range: ±15HL.