Instruction Manual 

FK Tensio 

 

FK_T-BA-e-1110 

 

2.  Working Conditions 

 
10°C to 30°C / 15% up to 80% humidity 

 

3. 

Electrical Power Supply 

 
Either by batteries or current power supply 
 
Power supply:  

Connection by power adapter 

 Batteries (6 x AA batteries

)

 

 

4. 

Technical Data 

 

Accuracy: ± 0,5 % of Capacity 

Data Sampling Rate: 1.000 Hz 

 

 

Model 

Capacity 

Resolution 

FK 10T 

10 N 

0,005 N 

FK 25T 

20 N 

0,010 N 

FK 50T 

50 N 

0,020 N 

FK 100T 

100 N 

0,050 N 

FK 250T 

200 N 

0,100 N 

 
 

5. 

Operation 

 

a. 

Display 

 

 

 
Measuring Result 
Measuring Units 
Measuring Direction 
Indication of PEAK Mode 
 

b. 

Operating keys 

 
ON / OFF:  
ON / OFF key 
(For ON, press 1 sec.) 
 
MAX:  
Changes between Track and Peak mode (capture of 
maximum values) 
 
UNIT:  
Press shortly:  

 

Select unit: 

N, kg, lb, ou 

ZERO:  
Zeros the measuring result (Tara function) 
Cleans the peak value (in Peak mode) 
 

c. 

Display Return 

 
When the instrument itself is turned   that the head faces 
down, the display returns automatically.  
 
 

d. 

 Measurement (Track Modus) 

 
Display (1) shows the continuous force in a defined 
direction (3)  
   
To zero the display, press ZERO:  
 

e. 

Peak-Hold Function (Peak Mode) 

 
Please press: MAX 
 
 
 
 
 

6. 

Backside Fixture to a test stand 

 

Fixture by 4 x M3 screws. 
 

7. 

Adjusting procedure 

  
Observe stable environmental conditions. A short warming 
up time of approx. 1 minute is recommended for 
stabilization.  
 
Switch on the unit by pressing the Unit-key and keep it 
pressed until in the display first appears shortly "CAL".  
 
After that there appears the fix  size of the adjustment 
weight. Attach the adjustment weight, a short time later 
appears "F". Then the  instrument returns automatically to 
the measuring mode. 
 
In case of an adjustment error or incorrect adjusting weight 
the display will show "E"; repeat adjustment process. 
 

8.  CE Declaration of Conformity  

 

(1) 

(2) 

(3) 

(4) 

Содержание FK Tensio

Страница 1: ...cell is inside the housing Table of contents 1 Included in delivery 2 Working conditions 3 Electrical power supply 4 Technical data 5 Operation 6 Backside fixing on a test stand 7 Adjusting procedure 8 Declaration of conformity 1 Included in delivery SAUTER FK Power cable Standard attachments as shown below Terms Track continuous measurement Peak Capture of the maximum value The measuring device o...

Страница 2: ...t unit N kg lb ou ZERO Zeros the measuring result Tara function Cleans the peak value in Peak mode c Display Return When the instrument itself is turned that the head faces down the display returns automatically d Measurement Track Modus Display 1 shows the continuous force in a defined direction 3 To zero the display press ZERO e Peak Hold Function Peak Mode Please press MAX 6 Backside Fixture to...

Страница 3: ...Structural alterations may not be made to the instrument This can lead to incorrect measuring results faults concerning safety regulations as well as to destruction of the instrument The instrument may only be used in compliance with the described guidelines Varying areas of application planned use must be approved by SAUTER in writing 9 3 Guarantee The guarantee is not valid following non observa...

Страница 4: ...110 4 They are not suitable for hanging from a mechanical hook e g a crane hook 9 5 1 Observe the information in the operating instructions please read the operating instructions carefully before erecting and commissioning even if you already have experience with SAUTER instruments 9 5 2 Staff training The device may only be operated and looked after by trained members of staff ...

Отзывы: