English
Quick start guide
Operating temperature, protection of the instruments and information about storage and the battery
1 Display
•
Operating temperature: 5 to 50 °C (KP 110-RF, KP 111-RF); -40 to 70 °C (KT 110-RF, KT 150-RF,
KTU 210-RF, KTT 310-RF, KTR 310-RF).
•
KP 110-RF, KP 111-RF, KT 110-RF, KT 150-RF, KTU 210-RF, KTR 310-RF: IP65. KTT 310-RF: IP54.
•
Storage temperature: from -40 to +85 °C
Battery power supply: lithium AA 3.6 V
2 Dimensions (mm)
Reference
Parameters
KP 110-RF / KP 111-RF
Pressure
KT 110-RF
Temperature / Current / Voltage / Pulses
KT 150-RF
Temperature
KTU 210-RF
Current / Voltage / Pulses / Light
KTR 310-RF
Pt100 temperature
KTT 310-RF
Thermocouple temperature
DATASET is finished
One value is being recorded
Slow flashing: dataset is taking 80-90% of storage capacity
Fast flashing: dataset is taking 90-100% of storage capacity
Constant: storage capacity filled up
Channel No. which is measuring
Refresh of displayed measurements
Displayed values correspond to maximum and minimum values of
the channels
Alarm action type: rising or falling action
REC
END
ACT
MIN
MAX
1
FULL
KP 110-RF
KT 110-RF
KP 110-RF / KT 150-RF:
KT 110-RF / KTU 210-RF
KTT 310-RF / KTR 310-RF:
Flashing: dataset has not started yet
Constant: data set is in progress
LOG
Display of measurement and recording intervals
TIME
Low battery indicator
Indicates the reception powerful of the datalogger
74.4
39.8
125.
8
74.4
116.
5
39.8
118.
6
74.4
39.8
KT 150-RF
KTR 310-RF / KTT 310-RF
74.4
116.5
39.8
KTU 210-RF
74.4
161.
7
39.8
118.6
39.8
74.4
3 Connections
KP 110-RF
Ribbed connectors
Mini-Din connector
KT 110-RF
Mini-Din connectors
KTU 210-RF
Connectors for thermocouple probes
Mini-Din connections for Pt100 probes
KTR 310-RF
KTT 310-RF
Activate or deactivate the keys during a measurement dataset:
• Press “Select” and “OK” at the same time for 5 s to deactivate both buttons: red LED
flashes twice quickly.
• Press “Select” and “OK” at the same time for 5 s to reactivate them: green LED
flashes twice quickly.
5 Recorder function
KISTOCK datalogger is supplied turned off to save battery life. To activate it:
• Press “OK” key for 5 s, green LED “On” blinks.
To turn it off:
• Press “Select” and “OK” keys at the same time for 5 s.
It is impossible to turn it off during a recording or a data download.
Perform an autozero during a measurement dataset (KP 110-RF)
It is possible to make the instrument to perform an autozero during a measurement dataset.
Keys must be activated.
• Disconnect the pressure tube from the instrument.
• Press “OK” key during 5 seconds.
The instrumetn performs the autozero: “0” is displayed on the screen, this value is not recorded.
• Connect again the pressure tubes.
The instrument starts again the measurements dataset and values are recorded.