62
K000000000EJ 032433 210519
On site calibration
Sensor CO
15/01/14
10:00
Is
uA
2.22
Measured gas
ppm
8000.0
Applied gas
ppm
8000.0
Elapsed time
00:03:00
Status
not active
Calibrate
Ia
uA
0.17
OK
OK
On site calibration
Sensor CO
15/01/14
10:00
Is
uA
2.22
Measured gas
ppm
8000.0
Applied gas
ppm
8000.0
Elapsed time
00:03:00
Status
active
Calibrate
Ia
uA
0.17
OK
→
Messages in the 'Status' line:
saving:
the instrument is saving the
performed calibration
error:
the sensor has NOT been
recalibrated for any of the following
reasons:
- The calibration gas cannot
properly reach the instrument.
- Concentration for the calibration
gas has not been set in the
relevant line ‘Applied gas’.
- The user didn’t allow for the
stabilization time to properly
elapse.
- The sensor could be damaged or
exhausted and must therefore be
replaced.
When the stabilization time is over, select the 'Calibration' row and activate the function ' ' to store the
new calibration.
OK
WARNING
At any time the user can restore the factory calibration in the instrument by setting
the 'Status' line on 'not active'.