6-6. Low Battery Warning Function 

The low battery indicator starts flashing when the battery power becomes low. If this occurs, immediately 

replace the batteries with new ones. Not doing so can affect the measurement accuracy or cause a 

malfunction 

 
 

7. Error Message 

If a problem is detected in the unit, an error code appears on the display. 

Error code

 

Description

 

Action

 

E_0 

Battery power is low.

 

Replace with new batteries.

 

E_4 

Internal processing error

 

- Remove the batteries, leave the unit for a 

while, reinstall the batteries, and then turn on 
the unit. 

- Replace with new batteries. 

 

E_5 

The sensor probe is not connected.

 

Securely connect the sensor probe.

 

E_8 

Sensor signal error

 

Hi 

The measured value is above the upper 
display limit.

 

Use the unit within the measurement range.

 

Lo 

The measured value is below the lower 
display limit.

 

If the displayed error code is not listed above or the problem persists after taking the recommended action, 
the unit or the sensor is likely to be faulty. 

Contact us or the store where you purchased the product.

 

 

 

8. Troubleshooting 

 

Problem

 

Possible cause

 

Action

 

The power cannot 
be turned on.

 

The batteries have run out. 

Replace with new batteries.

 

 
 
 
 
Readings are 
unstable.

 

The sensor connector is loosely 
connected. 

 

Correctly connect the sensor connector. 

The unit has not adapted to the 
ambient temperature.

 

Wait until the unit has adapted to the ambient 
temperature before starting measurement.

 

Insufficient quantity of the object to 
be measured or insufficient 
insertion depth of the sensor probe

 

Allow a sufficient insertion depth relative to 
the object to be measured 

 

The temperature of the object being 
measured varies widely.

 

Use the AVG function   

Affected by noise

 

Do not use the unit in a place where electrical 
noise or static electricity is generated.

 

 
Display error

 

The sensor connector is loosely 
connected 

 

Correctly connect the sensor connector 

 

Affected by noise

 

Do not use the unit in a place where electrical 
noise or static electricity is generated.

 

If the displayed error code is not listed above or the problem persists after taking the recommended 
action, the unit or the sensor is likely to be faulty. 

Contact us or the store where you purchased the 

product. 

Содержание SK-810PT

Страница 1: ...Instruction Manual Precision Digital Thermometer Model SK 810PT O 09...

Страница 2: ...t or near heating equipment Doing so may result in deformation of the casing or failure Do not use this unit in an environment where electrical noise is generated Doing so may result in unstable displ...

Страница 3: ...ve REL and moving average AVG values are calculated and displayed Hold function It is possible to freeze the display of measured values By using the HOLD function the measured values can be easily rea...

Страница 4: ...g displayed MAX Light up while the maximum measured value is being displayed AVG Light up while moving average value is being display or moving average value is being set REL Lights up while relative...

Страница 5: ...rtment The screws are integrated into the battery cover Do not forcibly pull them off as this could damage the cover 2 Install the batteries by referring to the polarity markings and provided on the b...

Страница 6: ...ove the cover before using the unit 5 3 Connecting the Sensor Probe 1 While the unit is off insert the sensor connector into the unit by aligning the concave part of the sensor connector with the conv...

Страница 7: ...lights up to indicate that the measured values are held 2 To cancel press the HOLD ENTER key again The HOLD indicator goes out and the unit returns to the normal temperature measurement mode The HOLD...

Страница 8: ...4 REL Function The REL function displays the relative values changes in measured values compared to the current measured value set to 0 00 C Operation procedure 1 Press the REL key during measurement...

Страница 9: ...the store where you purchased the product 8 Troubleshooting Problem Possible cause Action The power cannot be turned on The batteries have run out Replace with new batteries Readings are unstable The...

Страница 10: ...connected Accessories 4 AAA alkali batteries vinyl cover neck strap The specifications and the appearance subject to change with or without notice 10 Optional Items Products Cat No Descriptions Option...

Страница 11: ...pment The case that the products had been purchased in domestic and were sent to overseas by the purchaser is out of warranty Return of items Authorization must be obtained from us before returning it...

Страница 12: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 JAPAN TEL 81 0 3 3254 8117 FAX 81 0 3 3254 8123 URL http www sksato co jp english...

Отзывы: