SATO SK-810PT Скачать руководство пользователя страница 8

 

● Setting the measurement count 

(1) Press and hold the AVG key for at least two seconds. The AVG indicator lights up and the 

measurement count starts flashing. 

 

(2) Every time the AVG key is pressed, the measurement count changes.   
 

(3) Press the HOLD/ENTER key when the desired count is displayed. Confirm the set value and return 

to the normal temperature measurement mode. 

 
● Operation procedure 

(1) Press the AVG key during measurement. 

 

The AVG indicator starts flashing to indicate that the average value calculation is underway. During 

calculation, the measured values are displayed as bars. Upon completion of calculation, the AVG 

indicator changes from a flashing to solid light and the average value appears. 

 

(2) Pressing the AVG key again turns off the AVG indicator and returns to the normal temperature 

measurement mode. 

 

      - Pressing the HOLD/ENTER key while the average value is being displayed holds the display. 

 

6-4. REL Function 

The REL function displays the relative values (changes in measured values) compared to the current 

measured value (set to 0.00°C). 

 
● Operation procedure 

(1) Press the REL key during measurement. 

 

The REL indicator lights up and the measured value is displayed as "0.00°C". 

 

(2) Pressing the REL key again turns off the REL indicator and returns to the normal temperature 

measurement mode. 

 

      - Pressing the HOLD/ENTER key while the relative value is being displayed holds the display. 

 

6-5. Auto Power-Off Function 

The auto power-off function turns off the unit if there is no key operation for about 20 minutes. This saves 

the battery life, especially if you forget to turn off the unit.   

 
● Releasing the auto power-off function 

(1) When the unit is off, press and hold the POWER key for at least two seconds while pressing the 

HOLD/ENTER key. 

 

(2) When the "n" indicator appears on the display, release the keys. The "n" indicator goes out to 

indicate that the auto power-off has been released.   

 

Note: This function is no longer released once the unit is turned off. Perform the auto power-off release 

each time as necessary. 

 

Содержание SK-810PT

Страница 1: ...Instruction Manual Precision Digital Thermometer Model SK 810PT O 09...

Страница 2: ...t or near heating equipment Doing so may result in deformation of the casing or failure Do not use this unit in an environment where electrical noise is generated Doing so may result in unstable displ...

Страница 3: ...ve REL and moving average AVG values are calculated and displayed Hold function It is possible to freeze the display of measured values By using the HOLD function the measured values can be easily rea...

Страница 4: ...g displayed MAX Light up while the maximum measured value is being displayed AVG Light up while moving average value is being display or moving average value is being set REL Lights up while relative...

Страница 5: ...rtment The screws are integrated into the battery cover Do not forcibly pull them off as this could damage the cover 2 Install the batteries by referring to the polarity markings and provided on the b...

Страница 6: ...ove the cover before using the unit 5 3 Connecting the Sensor Probe 1 While the unit is off insert the sensor connector into the unit by aligning the concave part of the sensor connector with the conv...

Страница 7: ...lights up to indicate that the measured values are held 2 To cancel press the HOLD ENTER key again The HOLD indicator goes out and the unit returns to the normal temperature measurement mode The HOLD...

Страница 8: ...4 REL Function The REL function displays the relative values changes in measured values compared to the current measured value set to 0 00 C Operation procedure 1 Press the REL key during measurement...

Страница 9: ...the store where you purchased the product 8 Troubleshooting Problem Possible cause Action The power cannot be turned on The batteries have run out Replace with new batteries Readings are unstable The...

Страница 10: ...connected Accessories 4 AAA alkali batteries vinyl cover neck strap The specifications and the appearance subject to change with or without notice 10 Optional Items Products Cat No Descriptions Option...

Страница 11: ...pment The case that the products had been purchased in domestic and were sent to overseas by the purchaser is out of warranty Return of items Authorization must be obtained from us before returning it...

Страница 12: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 JAPAN TEL 81 0 3 3254 8117 FAX 81 0 3 3254 8123 URL http www sksato co jp english...

Отзывы: