5
Care after use
Thoroughly wash the sensor after use.
If the unit is stored with sample solution adhering to it, a coating could form. This coating will
adversely affect the measuring values or cause a malfunction; thus, make sure to wash the sensor
after each use.
Washing the sensor
T
he sensor must be washed to keep it clean and functioning properly.
Wet a soft sponge with commercially available ordinary neutral detergent. When washing the
sensor, do not use any special solutions (alkaline or acid) or abrasives.
Gently wash the sensor by rubbing it with the sponge, then rinse thoroughly with clean tap water.
It is recommended not to wipe the sensor after washing; leave it to air dry.
After use, protect the sensing part by covering it with
Sensor protective tube.
Specifications
Cat. No.
No. 1731-00 Digital Salt Meter
Model
SK-10S
Measuring factor
Salinity concentration
Detecting method
Electric resistance type
Measuring range
0.1 to 10.0% (at -5 to 70
℃)
Accuracy
±0.1 at 0 to 1%,
±
0.2 at 1% or more.
±
0.3 at 2% or more
±0.5 at 3% or more,
±
1.0 at 5% or more.
±
1.5 at 8% or more
Functions
Average, Measuring in interval, Displaying gram equivalent.
Auto power-off, Low battery
Power requirement
One 9VDC battery (006P)
Consumption current 150mA (Max.)
Operation ambient
0 to 55
℃
, 20 to 80%rh (no condensing)
Dimensions
Main body: 62 (w) x 148 (H) x 24 (D)mm
Probe: 198 (L) x 16 (W) x 16 (D)mm
Cord length: 0.5m
Weight
145g (including a battery)
Accessories
One 9V battery (6F22 006P), carrying case