    NB: If any other information appears at this point, it is likely that the sensing part is dirty. Wash the 

            tip of the sensor and wipe it off with a clean cloth or the like, and then press the Power/Measure 

            key again.   

            Washing the sensing part thoroughly before use will improve measurement accuracy. Always   

            wash the sensor before use especially when it is intended to be used for measuring high-salinity 

            solutions.

 

(Refer to "Washing the sensor".)

 

 

3. Immerse the sensing part into the sample, and then press the Measure key on the probe (or press   

    the Power/Measure key on the main unit). 

    NB: If the tip of the probe is not immersed in the sample deep enough, accurate measurement may 

            not be achieved.    Likewise, do not immerse the tip too deeply to avoid the possible   

            influence of the bottom of the container on the measurement values.

 

 

   

           

Use nonconductive containers such as glass, pottery and ceramic for accurate 

                  measurement. Do not use Metal containers. 

 

 

 

Calculating the average value

 

Stable measurement can be achieved by measuring several times and calculating the average value.

 

The unit calculates the average of up to nine measurements although it is possible to measure as many 

times as needed.

 

 

 

Press the Average key after at least two measurements (no operation when pressed after one 

measurement) to obtain the average value. 

 

The display starts flashing and the average value appears together with [AVE] on the display.   

        The numerical value on the upper-right corner indicates the number of measurements performed.   

        The average value is rounded off to one decimal place and displayed.

 

 

Auto power-off function 

The  unit  is  equipped  with  an  auto  power-off  function. 

This  function  turns  off  the  unit  if  there  is  no 

operation for about one minute. 

The range of concentration measurement and gram equivalent are retained.

 

        While measuring in intervals, the unit does not automatically turn off. 

 

 

Warning for Low Battery 

When the battery is exhausted, "LO-BAT" appears on the display. Replace the battery to new one.   

Otherwise, operation will not be possible. 