The Recall current configuration button enables the reading at any time of the
measurements recorded in the dosimeter.
The Apply new configuration button enables the new configuration to be validated and
the new parameters to be transferred to the dosimeter.
3.3.2
Configuring the Alarms
The Alarms tab enables the information relative to the alarms to be read and configured.
Figure 15 : Alarm configuration screen of the dosimeter
Ö
Threshold (V/m): in the drop-down menu choose the reference threshold to be used
by the device to trigger a visual alarm (20, 40, 60, 80, 100 or 140 V/m).
Ö
Buzzer: in the drop-down menu, choose the alarm function mode : Inactive,
Instantaneous or Mean 6 mn. If necessary, (in Instantaneous or Mean 6 mn mode)
establish the trigger value of the alarm (between 5 and 137V/m).
Ö
Vibrator: in the drop-down menu choose the alarm function mode : Inactive,
Instantaneous or Mean 6 mn. If necessary, (in Instantaneous or Mean 6 mn mode)
establish the trigger value of the alarm (between 5 and 137V/m).
Ö
Recall current configuration: this button enables the threshold, buzzer and vibrator
parameters to be reread at any moment.
Ö
Apply new configuration: this button enables the new configuration to be validated
and the new parameters to be transferred to the dosimeter.
3.3.3
Configuring Information
The Information tab is only displayed if the embedded software version of the device
allows this function. It enables the information relative to the measurement cycle to be
configured.
Figure 16 : Information configuration screen of the dosimeter