page 12
Manual SA 235
1. the display shows H1 = 1 point adjustment procedure is selected.
2. now place accurate weights of 1/3 of the maximum weighing range
on the weigh platform. (e.g. for a 3 kg scale: place 1 kg)
3. confirm with key , automatically followed by:
4. the weight adjustment will be executed, the most right display figure
runs down from 5 to 0 .
5. the display shows: H2
6. now place accurate weights of 2/3 of the maximum weighing range
on the weigh platform. (e.g. for a 3 kg scale: place 2 kg)
7. confirm with key , automatically followed by:
8. the weight adjustment will be executed, the most right display figure
runs down from 5 to 0 .
9. the display shows: H3
10. now place accurate weights of 3/3 of the maximum weighing range
on the weigh platform. (e.g. for a 3 kg scale: place 3 kg)
11. confirm with key , automatically followed by:
12. the weight adjustment will be executed, the most right display figure
runs down from 5 to 0 .
13. the scale turns off automatically, remove the weight.
14. the scales switches itself on again in the weighing mode.
After completion of the adjusting procedure it is advised to check the scale reading with accurate weights.
Repeat the adjustment procedure if necessary.
The weigh platform is overloaded, a warning "beep" sounds.
Remove the overload as soon as possible.
Weight reading is gross negative. Zero setting and taring is not possible.
The scale has been switched on with a load on the platform and has set
zero with this load.
After that the load has been removed and now the weight reading is negative.
Switch the scale off and on again with an empty weigh platform.
The scale refuses a zero setting key command.
The measured weight value is outside the zero range (+ and - 2% of the
weighing capacity of the scale).
Zero setting with the tare key is still possible If necessary switch the
scale off and on again with an empty weigh platform: a new zero setting
will be executed.
It is not possible to switch on the checkweigh- or countfuncion.
If one of these functions already is active, it is not possibly to switch on the
other one. Switch off the active function with key > 2 sec.
Switch the scale off and on again.
WEEGTECHNIEK HOLLAND B.V.
Website www.satex.nl
Tel +31 36 522 20 30
Patroonsweg 23
Email [email protected]
Fax +31 36 522 20 60
3892 DA ZEEWOLDE The Netherlands
Errors and reports
3 points adjustment: H1
Recommended for H2 and H5 modes of the weight reading
returns
from 5 to 0
returns
from 5 to 0
returns
from 5 to 0