Getting Started
§
Use the connecting cable from the weighing
platform to connect the display and control unit.
Note:
The cable gland is installed at the factory.
Please use extreme caution when performing any
work on the equipment that affects this cable gland.
Use a torque wrench
and tighten the cable gland to 5 Nm.
§
Prepare and install the cable as follows:
– Guide the cable through the cable gland.
– Close and tighten the cable gland in accordance
with the applicable standards.
– Remove the casing from a section of the cable end
(see illustration). The shield (1) must have contact
with the clamps (2).
– Expose approximately 15 cm (4 inches) of the wires
(3) for connection to the terminals.
– Guide the cable through the cable gland.
– It is important to make sure that the shield is in
contact with the clamps, because the cable is
grounded by the shield.
5