10
Weighing
The
R 160 P
has a second display range which you can easily
select at the touch of the soft key selector
(13)
.
lf you select the second range, only four decirnal places
(instead offive) will be displayed.
Standard display range: 0.00000 g
Second display range: 0.0000 g
The balance will stabilize at a faster rate in the display range
with an accuracy of 4 decimal places - 100 µ g readability - than
in the one with five decirnal places -10 µ g readability.
This selector provides you with a readily accessible weighing-in
convenience feature so that you do not have to access the
menu to change the codes set in the balance operating
program.