background image

  Leveling the Weighing Platform

 Purpose:

–  To compensate for uneven areas at the place of 

installation

–  To ensure that the equipment is placed in a perfectly 

horizontal position for consistently reproducible 

weighing results

  Always level the weighing platform again any time 

after it has been moved to a different location�

§

  Level the weighing platform using the four leveling 

feet� Turn the feet until the air bubble is centered  

in the level indicator�

§

  Check to ensure that all leveling feet rest securely  

on the work surface�

>  Each of the leveling feet must support an equal 

load�

>  Adjusting the leveling feet: 

To raise the weighing platform, extend the leveling 

feet (turn counterclockwise)� 

To lower the weighing platform, retract the leveling 

feet (turn clockwise)�

5

  For the User

–  Always make sure the equipment is  

disconnected from AC power before  

performing any installation, cleaning, 

maintenance or repair work on the scale�

–  If you see any indication that the scale 

cannot be operated safely (for example, 

due to damage), disconnect the scale 

from AC power and prevent use of the 

equipment for the time being� 

–  Chemicals (e�g� gases or liquids) that can 

corrode and damage the inside or outside 

of the device must be kept away from the 

equipment� 

  Handle the equipment and any accesso-

ries in accordance with the IP rating (IEC/

EN 60529)� 

–  The casing on all connecting cables, as 

well as the casing on wires inside the 

equipment housing, is made of PVC or 

rubber�  

–  The permitted ambient temperature range 

during operation is 0°C to +40°C�

  Make sure that the equipment is prop-

erly ventilated to prevent the buildup of 

excessive heat�

–  Only use original Sartorius spare parts�

–  Never use a hammer to close a paint can 

while is it still on the load plate, as this 

will damage the weighing system�

Содержание PMA.Power

Страница 1: ...1000025284 Operating Instructions Sartorius PMA Power Model PMA35001 Y Electronic Paint mixing Scale for Use in Zone 2 Hazardous Areas...

Страница 2: ...mixing applications 9 Mounting bracket for display and control unit 10 Interface 25 contact D Sub socket 11 Grounding terminal 12 Power cord with plug 13 Level indicator 14 Cable connection 15 Load p...

Страница 3: ...g the equipment in a Zone 2 hazardous area make sure you observe the warnings and safety precautions in their entirety during installation and oper ation as well as while performing mainte nance and r...

Страница 4: ...ce con tact your Sartorius dealer or the Sartorius Service Connect all equipment including periph eral devices to the equipotential bonding conductor The display values can be impacted when subjected...

Страница 5: ...sure the equipment is disconnected from AC power before performing any installation cleaning maintenance or repair work on the scale If you see any indication that the scale cannot be operated safely...

Страница 6: ...r scale will not be exposed to drafts heat radiation moisture or vibrations Make sure that you read these operating instructions carefully before connecting the scale to the power supply Observe all w...

Страница 7: ...y Information documents Plug the power cord into a wall outlet mains Observe all warnings and safety precautions See also Safety Instructions documents Install the scale in such a way or protect the p...

Страница 8: ...is displayed Add the other components up to the desired weight formula Remove the filled paint can from the load plate Never use a hammer to close a paint can while is it still on the load plate as th...

Страница 9: ...or 0 01 increments up to a factor of 1 0 Note A flashing arrow n on the display indicates that the weight value displayed is not valid in legal metrology i e not legal for trade Example As you pour th...

Страница 10: ...erpour occurred After the correction has been completed you can continue filling the remaining components Note You can correct overpours as often as needed Keep in mind that the total quantity of pain...

Страница 11: ...8 COR 01 11 The readout shows the amount of the first compo nent to be added and C1 is shown in the upper right hand corner of the display 1 7 g 12 Add until value reaches 0 0 g 0 0 g 13 Press the L k...

Страница 12: ...6 g Always allow approximately 30 minutes for the scale to warm up after connecting it to AC power and before performing calibration adjustment Press the U key Zero tare 3 for 2 seconds 10000 is displ...

Страница 13: ...desired menu item on this level Press the L key ENTER to select the next menu level The third menu level is displayed Use the RS keys to select the desired menu item on this level Press the L key ENT...

Страница 14: ...Level 3 Level 4 SETUP Press the L key ENTER SCALE Press the L key ENTER UNIT Use the RS keys to select the desired setting e g UNIT o GRAMS Press the L key ENTER PT PD Use the RS keys to select the de...

Страница 15: ...active setting Press the c key Clear Activating the LOCK Function The LOCK function protects the scale from unauthorized use When this function is active the scale readout shows weight values only wh...

Страница 16: ...f your password in a safe place The LOCK function can be deactivated only with this password Level 1 Level 2 Level 3 INPUT Use the RS keys to select INPUT PASSWORD Press the L key ENTER Press the L ke...

Страница 17: ...ord protected Level 1 Level 2 Level 3 Level 4 SETUP Press the L key ENTER RESET Use the RS keys to select RESET Press the L key ENTER MENU Use the RS keys to select MENU Press the L key ENTER YES Use...

Страница 18: ...he scale in another changes constantly area Too much vibration or the Access the menu to select scale is exposed to a draft the appropriate code for the ambient conditions see Menu Settings The weight...

Страница 19: ...lease refer to the information under Safety Inspection below Safety Inspection Safe operation of the scale is no longer ensured in the following cases There is visible damage to the AC adapter power s...

Страница 20: ...rning Risk of injury due to contaminated devices Devices contaminated with hazardous materials NBC contamination will not be accepted by Sartorius for repair or disposal Dispose of Device and Parts In...

Страница 21: ...it data Pin 4 DTR Data terminal ready Pin 5 GND Ground Pin 6 BPI switch no connection to the PMA35001 Y Pin 8 CTS Clear to Send Installing the Adapter Cable The adapter cable is required for operation...

Страница 22: ...del modelo S N Made in Germany II 3G Ex nA ic IIB T4 Gc DEKRA 19ATEX0006X IECEx DEK 19 0004X PCEC CE19 2nnnX Siehe Sicherheitshinweise See Safety Instructions Voir les consignes de s curit Vedi le ist...

Страница 23: ...N60529 Ambient conditions Indoor use only Height up to 2 000 m Supply voltage fluctuations up to 10 of the supply voltage Overvoltage Category II Pollution Degree 2 Load plate dimensions mm 350 240 Sc...

Страница 24: ...Mains cable Use only cable delivered by Sartorius 4 3 15 3 8 Hazardous Area Zone 2 Group IIB T4 Ambient temperature range 0 C 40 C Non Hazardous Area Non Hazardous Area or Hazardous Area Zone 2 Group...

Страница 25: ...in The cable connections in the Ex area must be secured against self loosening and the screw connections must be screwed tight Third party cables are the responsibility of the operator WARNING DO NOT...

Страница 26: ...26...

Страница 27: ...27...

Страница 28: ...28...

Страница 29: ...29 2019 03 05...

Страница 30: ...30...

Страница 31: ...31...

Страница 32: ...feminine forms are used to facilitate legibility in these instructions and always simultaneously denote the other gender as well Copyright notice This instruction manual including all of its component...

Отзывы: