Error Codes
Error Codes
Problem
Cause
Remedy
No segments appear
on the weight display
– No AC power is
available
– Check the AC power
supply
The weight readout
shows “Low"
– No load plate on the
scale
– Position the load plate
The weight readout
shows “High"
– Weighing capacity
exceeded
– Unload the scale
The weight readout
changes constantly
– The setup location
is unstable
– Excessive vibration
or draft
– Change the setup location
– Make the necessary
adjustment via the scale
operating menu
(see “Menu Settings")
The weight readout is
obviously incorrect
– The sample is not
stable
– Scale not tared before
weighing
– Tare before weighing
No weight value is
shown and the padlock
symbol
is displayed
– Communication
between scale and PC
has been interrupted,
activating the LOCK
function
– Access the menu to make
the necessary adjustment
and switch off the LOCK
function
– Check the connection
Содержание PMA.Power Series
Страница 29: ...Operating Instructions PMA 35001 X 29 EC Declaration of Conformity...
Страница 30: ...30 Operating Instructions PMA 35001 X ECDeclaration of Conformity...
Страница 32: ...32 Operating Instructions PMA 35001 X EC Verification...
Страница 33: ...Operating Instructions PMA 35001 X 33 EC Verification...
Страница 34: ...34 Operating Instructions PMA 35001 X EC Verification...
Страница 40: ...40 Operating Instructions PMA 35001 X Certificate of Compliance...
Страница 41: ...Operating Instructions PMA 35001 X 41 Certificate of Compliance...
Страница 42: ...42 Operating Instructions PMA 35001 X Certificate of Compliance...
Страница 43: ...Operating Instructions PMA 35001 X 43 Certificate of Compliance...