19 

Troubleshooting Guide 

 
 
 
 
 
 

Problem ... 

Causes ... 

Remedy 

 

No segments appear in the weight 
display 

(16)

 

 

No line voltage available 

The AC adapter has not been 

plugged in 

 

Check current supply 

Plug in the AC adapter 

 

Weight display shows 

“L”

 or 

“CH2” 

 

The pan 

(1)

 is not in place 

 

Place the pan 

Weight display shows 

“H” 

Load exceeds capacity of the 

balance 

Unload the balance 

 

The weight readout changes con-
stantly or the special message 

“BUSY”

 does not go out in the 

weight display 

 

Unstable ambient conditions 

Too much vibration or there is a 

draft 

 

The door of the weighing cham-

ber is not completely closed (for 
B 120S) 

 

The sample or object does not 

have a stable weight 

 

Set up scale in another area 

Access the menu to adapt scale 

to the particular type of weigh-
ing environment 

Close the door of the weighing 

chamber 

 

The weight display shows 

“CE” 

 

The CAL key 

(14)

 was not 

pressed when the display read 
zero 

The balance is loaded 

 

Press tare control 

(10)

 and re-

press the CAL key 

 

Unload the balance 

 

 

The weight readout is obviously 
wrong 

 

The balance has not been cali-

brated 

The balance has not been tared 

before weighing 

Scale has not been tared before 

weighing 

The air bubble of the level indi-

cator 

(9)

 is not within the circle 

(only for B 120S, B 210P/S,  
B 3100P/S) 

 

Calibrate the balance 

 

Tare before weighing 

 
 
 

Level the balance 

 
 
 
 

Содержание Master series

Страница 1: ...Sartorius Basic Series Electronic Analytical and Precision Balances Installation and Operating Instructions...

Страница 2: ...2 B 120S...

Страница 3: ...3 B 310P B 310S...

Страница 4: ...4 B 410 B 610 B 3100P B 3100S B 4100 B 6100...

Страница 5: ...cylinder only B 310P and B 310S 7 Pan adapter only B 310P and B 310S 8 Power receptacle 9 Level indicator only B 120S B 310P B 310S B 3100P and B 3100S 10 Leveling foot only B 120S B 310P B 310S B 310...

Страница 6: ...ipment delivered Electronic Analytical or Precision Balance of the Basic Series is manufactured and tested in compliance with the following DIN VDE regulations DIN IEC 348 VDE0411 Safety requirements...

Страница 7: ...ns 9 How to Start Up the B 120 S 10 How to Start Up the B 310P and B 310S 11 How to Start Up the B 410 B 610 12 B 3100P B 3100S B 4100 and B 6100 Operation 13 Calibration 15 Balance Operating Program...

Страница 8: ...ance AC adapter Dust cover Pan adapter Base plate of the round draft shield Weighing pan Glass draft shield cylinder Round draft shield lid Balance AC adapter Dust cover Weighing pan Do not miss out o...

Страница 9: ...apt it to your particular requirements and to the specific ambient condi tions simply by changing the menu code settings in the balance operating program For this purpose please read pages 16 to 18 Af...

Страница 10: ...ting printed on this unit is identical to your local line voltage rating Plug the cord of the AC adapter into the power receptacle Se cure the connection by tightening the knurled collar Now plug the...

Страница 11: ...s powered by an AC adapter Make sure that the voltage rating printed on this unit is identical to your local line voltage rating Plug the cord of the AG adapter into the power receptacle of the balanc...

Страница 12: ...n this unit is identical to your local line voltage rating Plug the cord of the AC adapter into the power receptacle of the balance For models B 410 and B 610 secure the connection by tightening the k...

Страница 13: ...y to operate without warmup POWER OFF The balance was disconnected from line power reconnection to line power after the balance was unplugged power failure or outage CAL The calibration function has b...

Страница 14: ...power is turned on an automatic self test of all essen tial electronic functions is performed This self test ends with one of the following readouts 0 0000 g 0 000 g 0 00 g 0 0 g one of these readouts...

Страница 15: ...econds until the calibration weight readout appears in the display Center the calibration weight on the weighing pan Now the weight unit symbol is displayed An acoustic Signal in dicates the end of th...

Страница 16: ...urned off STANDBY state hold down the tare control 13 while you press the ON OFF key 15 Upon completion of the automatic self test release the tare control as soon as CH5 is displayed The Status of th...

Страница 17: ...e The o symbol that appears indicates the actual setting To change any menu code settings C mode press the tare control as soon as the appropriate code is displayed Brief display of BUSY and the o sym...

Страница 18: ...rmat c 5 2 2 Singapore taels tl C 1 3 1 Last decimal ON c 5 2 3 Taiwan taels tl C 1 3 2 Last decimal OFF c 5 2 4 Grains gr C 1 3 3 Last decimal at stability c 5 2 5 Pennyweights dwt C 1 3 4 All decima...

Страница 19: ...g cham ber is not completely closed for B 120S The sample or object does not have a stable weight Set up scale in another area Access the menu to adapt scale to the particular type of weigh ing enviro...

Страница 20: ...x lines sec 1 5 Printer housing W x D x H in mm 150 x 138 x 43 in inches5 9 x 5 4 x 1 7 Data printer with set of programs 73802 to order replace with one of the codes below 201 for animal weighing 207...

Страница 21: ...tery pack YRB 01Z approx 12 hours of Operation rechargeable via AC adapter only for B 3100S P B 4100 B 6100 Carrying case forB120S YDB 01A for B 310 P S YDB 01L for B 410 B 610 YDB 02L B 3100 P S B 41...

Страница 22: ...283 303 10 o C 30 o C 273 313 0 o C 40 o C Allowable relative humidity 25 85 15 85 moisture proof rating Sensitivity drift within 10 30 o C o C 2 10 6 5 10 6 2 10 6 Pan size mm in 90 3 5 130 5 1 130 5...

Страница 23: ...bility range d 0 25 64 selectable Ambient temperature range K 273 313 0 o C 40 o C Allowable relative humidity 15 85 moisture proof rating Sensitivity drift within 10 30 o C o C 5 10 6 Pan size mm in...

Страница 24: ...ty range d 0 25 64 selectable Ambient temperature range K 273 313 0 o C 40 o C Allowable relative humidity 15 85 moisture proof rating Sensitivity drift within 10 30 o C o C 2 10 6 5 10 6 5 10 6 Pan s...

Страница 25: ......

Страница 26: ...Copyright by Sartorius AG G ttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartori...

Отзывы: