![Sartorius IW2P1x-150 series Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/sartorius/iw2p1x-150-series/iw2p1x-150-series_operating-instructions-manual_1199745019.webp)
Miras 2
Miras 2
19
19
Check Weighing
Step
Key (or instruction)
Display
a. Select application program
(Fn/Up) > 2 sec
b. Select “Checkweighing”
(Fn/Up) or (G/N/Down)
repeatedly
c. Press Print/Enter to confirm.
(Print/Enter)
“LO” flashes.
d. Set lower limit
(Fn/Up) or (G/N/Down) and/or
(Zero/Left) or (Tare/Right)
e. Press Print/Enter to confirm
(Print/Enter)
and save. “HI” flashes
f. Set upper limit,
(G/N/Down) and/or
and then press Print/Enter.
(Zero/Left) or (Tare/Right)
g. Press Print/Enter to confirm
(Print/Enter)
and save.
h. Place the sample on the scale.
Display shows the difference
and HI/LO symbol
i. Unload the scale
j. Checkweighing application: Clear (Zero/Left) > 2 sec
Содержание IW2P1x-150 series