44 Entris
®
II Advanced Line Operating Instructions
Malfunctions
10 .2 Troubleshooting
Fault
Cause
Remedy
Chapter,
page
The operating display is
blank .
The device is disconnected .
Check the connection to the power
supply .
The power supply unit is
not
connected .
Connect the power supply cable to
the power supply .
The displayed weight
value changes
constantly .
The installation site is unstable .
Adjust the parameter for the
ambient temperatures .
Change the installation site .
A foreign object is positioned
between the weighing pan and
the housing .
Remove the foreign object .
The weight readout
displayed by the device
is obviously wrong .
The device has
not
been
calibrated and adjusted .
Calibrate and adjust the device .
The device was
not
tared before
weighing .
Tare the device .
For a conformity-
assessed device: The
[
Invalid
weight value]
display appears .
The cause of this fault is
displayed after pressing the [
No
valid weight value] button .