Malfunctions
Warning message
Fault
Cause
Remedy
Chapter,
page
ERR 10
Taring is
not
possible .
The device
cannot
be
manually tared because
an application program
has the tare memory
reserved .
To release the tare memory:
Exit the application
program with the [End]
button .
ERR 11
The weight value
cannot
be saved in
the tare memory .
The weight value is
negative or “zero” .
Check the sample being
weighed .
Zero the device before
placing the sample on the
balance .
Low voltage of
battery of the
internal clock
module: The date
and time may be
incorrect .
The battery is nearly
empty .
Contact Sartorius Service .
Calibration
canceled
The device
cannot
be
adjusted when
starting the
calibration function
due to a zero point
error or an absence of
leveling (conformity-
assessed models) .
The device was
not
zeroed before calibration .
Zero the device .
Check the preload and set if
necessary .
The device is loaded .
Remove the sample from
the weighing pan .
The balance drifts too far .
Allow to warm up . Re-align
the device .
The internal motor weight
unit is defective .
Contact Sartorius Service .
Conformity-assessed
models: The device is not
leveled .
The balance
needs to be
adjusted!
The device needs to be
calibrated and adjusted .
Calibrate and adjust the
device .
Remaining time in
minutes: xx
The device is in the warm-
up phase and has
not
yet
reached its operating
temperature .
xx = remaining time in
minutes
Comply with the warm-up
time after switching the
device on .
Weight value is
invalid
The displayed value is
invalid .
Zero the device .
The balance
needs to be
leveled!
The device is not leveled .
Level the device .