![Sartorius CPA 10001 Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/sartorius/cpa-10001/cpa-10001_operating-instructions-manual_1199719057.webp)
Data output example: +123.56 g
Position
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
+
1
2
3
.
5
6
g
CR
LF
+
1
2
3
.
5
[
6
]*
g
CR
LF
Position 1:
Plus or minus sign or space
Position 2:
Space or brackets*
Positions 3–10:
Weight with a decimal point; leading zeros = space
Position 11:
Space or brackets*
Positions 12–14: Unit symbol or space
Position 15:
Carriage return
Position 16:
Line feed
Output Format with 22 Characters
When data is output with an ID code, the ID code (consisting of 6 characters) precedes
the 16-character string described above.
These 6 characters identify the subsequent value.
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22
I
I
I
I
I
I
+
*
D
D
D
D
D
D
D
D
*
U
U
U CR LF
*
*
*
*
*
–
.
.
.
.
.
.
.
.
*
*
*
*
*
*
*
*
*
*
*
*
I:
ID code character
U:
Unit symbol
1
)
*:
Space
CR: Carriage return
D:
Digit or letter
LF: Line feed
Example:
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22
N
+
1
2
3
.
5
6
g
CR LF
N
+
1
2
3
.
5
[
6
]
g
CR LF
1
)
Depends on balance type; for example, not all units or characters are available on balances verified for use in legal
metrology.
*)
Identification of non-verified digits:
Non-verified digits where e
=
d are identified by brackets (square) if you select the following setting:
Parameter settings: Communication mode: Universal printer (menu code 5 5 2); “SBI” menu setting
In the “SBI” setting (menu code 5 5 1), non-verified digits displayed are not automatically identified as such.
Be sure to take the steps or make the settings on the auxiliary device for this purpose.
57