ISO/GLP-compliant Printout/Record
53
Features
You can have the parameters pertaining
to the ambient weighing conditions
printed before (GLP header) and after
(GLP footer) the values of a weighing
series. These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance manufacturer
– Balance model
– Balance serial number
– Software version number
– Identification number of the current
sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
The record is output to a Sartorius data
printer or a computer.
Settings
G
Setting menu codes for the printout
(see “Configuration"):
– ISO/GLP-compliant record after
calibration/adjustment only: menu
code
8 10 2
; or ISO/GLP-compliant
record always on: menu code
8 10 3
– Line format for printout:
With data ID codes - 22 characters:
menu code
7 2 2
!
No ISO/GLP-compliant record is output
if any of the following settings are
configured: menu code
6 1 4
,
6 1 5
,
6 1 6
(automatic printout) or
7 2 1
Function Keys
Press (
A
) to output header and first
measured value.
> Header is output the first time (
A
)
is pressed
Press (F) to include output header and
reference data on automatic printouts
when an application program is active
Close the application:
1) Output GLP footer:
Press (CF)
2) End application program:
Press (CF)