17
Example 1:
Count parts into a container
Menu settings: factory settings
Step
Press key
Display/Printout
1. Turn on scale, if power is off.
[ON/OFF]
2. Place empty container
[START]
0.0
g
on scale and start counting.
+
36.6 SAMPLE
3. Select reference sample quantity;
Brief display of:
for example, turn selector to
“20”.
ref 20
4. Place reference sample quantity
on scale (20 parts in this case).
+
23.1
g
5. Save reference sample quantity.
[OK]
+
20
pcs
The scale now calculates the
+
23.1
+
1.1550
average piece weight.
6. If necessary, update the reference
sample using the optimizing function;
+
32
pcs
in this case, add 2 to 20 additional
+
36.6
opt.
parts to the weighing pan.
7. Now place unknown number of parts
+
260
pcs
on pan (in this example, 260).
+
297,4
+
1.1438
8. Print piece count.
nRef
+
32
If you have selected
[PRINT]
wRef
+
1.1438 g
“nRef” and “wRef” in the menu,
Qnt+
260 pcs
these lines will be printed:
G
+
297.4 g
9. Unload scale
0
pcs
10. Repeat counting starting from step 7, if desired.