23
Example
Determine the weight of a sample under extremely unstable ambient conditions
by calculating the average of 20 subweighing operations.
Settings (changes in the factory settings required for this example):
Menu: “Weigh averaging” application program, code
2.1.12
Step
Press key
Display/Printout
1. Turn on scale, if power is off.
[ON/OFF]
2. Tare scale
[TARE]
0.0
g
3. Select number of subweighing
operations
ref 20
(briefly)
(20 measurements in this example)
4. Place sample on scale
+
8888
(weight readout fluctuates;
here, for example, by about 275 g)
5. Start measurement
[OK]
+
8888
20
1
9
...
1
After 20 subweighing operations
+
275.5
g
G
If the print format is set to include
data ID codes, the following is printed:
RES + 275.5 g
6. Unload scale
+
275.5
g
G
(stable display)
7. Delete result
[DELETE]
8. Repeat procedure starting from step 4, if desired.