13
Start up of cassettes
Sartoflow
®
10 | Sartoflow
®
20 and Sartoflow
®
Systems
with Hydraulic Piston, 12.5 cm
2
Cassette type
Order
number
Required
Compression
Sterilization* Hydraulic
Pressure
Sterilization*
Operating Conditions
Operating Conditions
Sartocube
®
and
Sartocon
®
Cassettes
302146... |
302154…
18–35 kN
< 10 kN
170 bar
(2466 PSI)
(17 MPa)
80 bar
(1160 PSI)
(8 MPa)
PESUmax
302146AL… 21–35 kN
No thermal
sterilization
170 bar
(2466 PSI)
(17 MPa)
No thermal
sterilization
Sartocube
®
and
Sartocon
®
Hydrosart
®
MF | UF
302144… |
302186…
18–25 kN
< 10 kN
170 bar
(2466 PSI)
(17 MPa)
80 bar
(1160 PSI)
(8 MPa)
* List of sterilizable cassettes is available in “Technical Specifications”, section 6.1.
** After every tenth time of use, the PEEK washers should be replaced. The bronze nut or
the trapezoidal thread should be replaced as soon as abrasion is visible. It is essential that
the bronze nut and the trapezoidal thread are kept clean.
5.3 Rinsing Cassettes
Flush the filter cassettes as indicated below:
− Before initial use with purified water
− With buffer or physiological saline before filtration
− With buffer or physiological saline before cleaning
− With purified water after cleaning
− With purified water after disinfection
− With purified water before checking the functionality of the filter cassettes
− With purified water before sterilization
Flush Procedure for Cassettes
The installed cassettes must be flushed with purified water or RO water at the recom-
mended approx. ratio of retentate to permeate flow as listed below.
Cassette Type
Ratio of Retentate to
Permeate flow
Ultrafilters 1 kD | 2 kD
Approx. 5 : 1
Ultrafilters 5 kD
Approx. 3 : 1
Ultrafilters 8 kD up to 300 kD
Approx. 1 : 1
Microfilters 0.1 µm
Approx. 1 : 1
Microfilters 0.2 µm and 0.45 µm and PESU MAX
Approx. 1 : 2
The cassette must be flushed with 10 l purified of water per 0.1 m
2
filter area.
5.4 Integrity Test
For multiple cassette integrity testing see ”#1 Technical Data Package Integrity testing
Sartocon
®
Cassettes“.
The Technical Data Package Integrity testing is available upon request from your local
application specialists.
Ensure the filter cassettes are sufficiently wet before measuring the air diffusion.