46 Appendix
Warning messages
Cause
Elimination
There are no valid parameters on this
machine . Please load parameters .
There are no parameters on the device .
Switch machine on and off .
If there is still no current data,
contact your Sartorius Stedim
Biotech Service Representative .
SD missing . Please insert SD and retry!
No SD card inserted
Insert card in SD slot and repeat the
action again .
SD not valid
Wrong SD card inserted
Use the SD card supplied by Sartorius
Stedim Biotech .
Login failed
Wrong user name or password entered
Enter correct user name with the
associated password
Insert validation holder
1 . Validation holder not inserted
2 . Validation holder inserted, identification
defective
1 . Insert validation holder
2 . Contact Sartorius Stedim Biotech
Service Representative .
Remove validation holder
1 . Validation holder was not removed
2 . Validation holder was removed, identifica-
tion defective
1 . Remove validation holder
2 . Contact your Sartorius Stedim
Biotech Service Representative .
Run stopped by user
User interrupts a sequence
Confirm message with OK
Log transfer stopped by user
Abort by the user while data being written to
the SD card
Confirm message with OK