![SARK 110 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/sark/110/110_user-manual_1199282048.webp)
SARK
SARK
SARK
SARK-
-
-
-110
110
110
110
User’s Manual
Rev 1.2.6 October 3
rd
, 2015
- 48 -
© Melchor Varela – EA4FRB 2011-2015
The following parameters should be supplied:
•
Sweep:
o
None
Continuous frequency mode
o
Frequency
Sweep frequency mode
•
Repeat:
o
Continuous
Continuous signal generation
o
Single
Signal generator stops after a single sweep
o
Count:
Signal generator stops when number of sweeps reach count
<Count>
•
Function:
o
Linear
Linear frequency increase or decrease
o
Log
Logarithmic frequency increase or decrease
o
Bi-Linear
Start-Stop-Stop-Start sweep (Linear)
o
Bi-Log
Start-Stop-Stop-Start sweep (Log)
•
Start Frequency:
o
<Start>
Hertz
•
Stop Frequency:
o
<Stop>
Hertz
•
Number of points:
Number of steps between start and stop frequency
o
<Points>
•
Delay uS:
Step time
o
<Delay uS>
Micro-seconds