The  remaining  capacity  is  15  ...  25%  depending  on  the  chemistry  of  the  used  cells.  After 
pressing  the  Toggle  button  display  presents  measurement  results  as  usually.  If  longer 
measurements  are  planned,  the  batteries  should  be  replaced  before  to  avoid  unexpected 
measurement break. If the cell voltage drops further, below 0.9 Volts, the running sample is 
stopped automatically and the device enters the stand-by mode with “CHANGE BATTERIES” 
indication. 
 

CHANGE BATTERIES     
    Thoron-Scout     
      SN:00256       
    CHANGE BATTERIES 

 
After  plugging  the  power  supply  and  waiting  one  minute  the  measurement  can  be  started 
anew. 
 
The uptake of power during stand-by is about 15 – 20 % of a measurement. Anyway: In case 
of storage of the instrument without usage, batteries should be removed. 
 

Carrying out a measurement 

Press short the Toggle button to start a new measurement series. The display will show the 
remaining time to complete the first integration interval. 
 

    Thoron-Scout     
      SN:00256       
  Wait 120 Minutes   

  for first data !     

 
The  actual  status  and  set-up  information  (see  below)  may  be  displayed  by  pressing  the 
button again. 
If the first interval has been completed, five different display pages are available. The several 
pages can be toggled by repeated pressing of the Toggle button. Depending on the selected 
system  of  units,  the  concentration  values  are  given  either  in  ‘Bq/m³’  or  ‘pCi/L’  (mbar/inHg, 
°C/°F) 
The  first  page  shows  the  actual  Radon  concentration  (calculated  for  the  last  sampling 
interval)  with  the  statistical  error  for  a  1  Sigma  confidence  interval.  If  the  “Fast”  mode  was 
selected,  a  starlet  is  appended  to  the  word  “Radon”  in  the  first  row.  Right  beside,  the  time 
stamp is given when the integration interval of the calculated concentration was finished 
The third row contains at the left hand side the total number of integration intervals since the 
last start of a measurement series. At the right hand side the pre-set integration interval and 
the remaining time period of the actual sample is displayed. 

 
  Radon*     12:20   

       85Bq/m³

±

10%  

  #34   117/120Min   
      TSC-00256      

 
Page two gives the same information for Thoron (Rn-220) 
 

  Thoron     12:20   

      124Bq/m³

±

16%   

  #34   117/120Min   
      TSC-00256      

 

Содержание Thoron-Scout

Страница 1: ...Manual_ThoronScout_EN_02 07 15 doc 1 User Manual Thoron Scout Version 06 2015 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMAN...

Страница 2: ...5 CARRYING OUT A MEASUREMENT 6 Adjustment of the sampling interval 7 OPERATION MODES 7 Pump 8 Alarm 8 Fast Slow Mode 8 Sniffing 8 DATA HANDLING 8 Data Storage 8 Data transfer RS232 and USB interface 8...

Страница 3: ...Minutes This time span defines the minimum achievable response time to a Radon concentration step Now the decay chain is continued by the both beta emitters Pb 214 and Bi 214 followed by another alph...

Страница 4: ...tes dislocation during the measurement A display with backlight informs about the actual readings Neither mechanical parts like membrane pumps nor an external power supply are required Therefore use a...

Страница 5: ...is placed at the rear panel of the Thoron Scout Changing the batteries will force you to adjust the re set internal real time clock A concerning message appears on the display of the Thoron Scout Sto...

Страница 6: ...oron Scout SN 00256 Wait 120 Minutes for first data The actual status and set up information see below may be displayed by pressing the button again If the first interval has been completed five diffe...

Страница 7: ...6 19 12 34 ALM 10000Bq m INTVL BUZZ OFF TSC 00256 To finish a measurement series keep the Toggle button down and wait for at least four beeps from the buzzer If the button is released the sample will...

Страница 8: ...should be cleared after successful data transmission and storage on hard disk This will save time during the next transmission and avoids redundant data storage Each data record is stored after compl...

Страница 9: ...e accuracy of the measurement From the opposite point of view one could ask How many counts I have to detect to achieve a predefined uncertainty Two items will affect the number of counted decays The...

Страница 10: ...her than the lower limit of the error band of reading 1 Because the true value could be placed within 1350 Bq m and 1469 Bq m the variation of both readings is not statistical significant Example 2 Re...

Страница 11: ...han 16 and therefore the statistical fluctuations have to be derived by the Poisson distribution The stated confidence interval gives the probability that the detected number of decays within the inte...

Страница 12: ...erval 1 255 Minutes adjustable Memory 2047 data records Internal volume chamber approx 60 ml Power supply Battery operation 30 days AC DC adapter 4 5V 0 5A PC Interface Serial USB or RS232 9600 Baud 8...

Отзывы: