Operating the Instrument 

General 

 

The new Thoron-Scout is a versatile, easy to use and state-of-the-art instrument focussed on 
the detection of Thoron (Rn-220) and Radon (Rn-222) in the ambient air. The required fast 
exchange rate of sampled air is realized by a highly permeable chamber placed outside the 
instruments  enclosure.  The  relative  Thoron  sensitivity  is  comparable  with  the  one  of  pump 
based instruments. 
 
Beside the activity concentration of Thoron and Radon, air temperature, relative humidity and 
barometric pressure will be determined and saved to a non-volatile circular memory (first in –
first out). Up to 2000 chronological data sets will be available for data transfer to PC including 
alpha  spectrums.  An  internal  real  time  clock  ensures  a  correct  time  regime,  a  tamper  lock 
indicates  dislocation  during  the  measurement.  A  display  with  backlight  informs  about  the 
actual readings. 
 
Neither mechanical parts like membrane pumps nor an external power supply are required. 
Therefore, use and exposition at home or at workplace is possible without any disturbance. A 
total duration of measurement of up to four weeks with continuous data recording is possible. 
 
Due to its small dimensions and little weight, the Thoron-Scout can be shipped by mail to the 
place  of  interest/measurement  without  any  additional  man-power  required  for  installation  or 
start-up of an analysis – even untrained staff is able to start a measurement. 
 
Entire part of delivery will be comfortable software for data read out and graph presentation, 
dose assessment and easy data backup. 
 

  Read out of measurement data and adjustment of device parameter  

  Interactive graphic display with zoom and pan  

  Automatic backup of measurement data  

  Selective export and conversion of customised time periods to text files for additional 

tasks (like import to EXCEL)  

 

Front panel elements 

 

 

Содержание Thoron-Scout

Страница 1: ...Manual_ThoronScout_EN_02 07 15 doc 1 User Manual Thoron Scout Version 06 2015 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMAN...

Страница 2: ...5 CARRYING OUT A MEASUREMENT 6 Adjustment of the sampling interval 7 OPERATION MODES 7 Pump 8 Alarm 8 Fast Slow Mode 8 Sniffing 8 DATA HANDLING 8 Data Storage 8 Data transfer RS232 and USB interface 8...

Страница 3: ...Minutes This time span defines the minimum achievable response time to a Radon concentration step Now the decay chain is continued by the both beta emitters Pb 214 and Bi 214 followed by another alph...

Страница 4: ...tes dislocation during the measurement A display with backlight informs about the actual readings Neither mechanical parts like membrane pumps nor an external power supply are required Therefore use a...

Страница 5: ...is placed at the rear panel of the Thoron Scout Changing the batteries will force you to adjust the re set internal real time clock A concerning message appears on the display of the Thoron Scout Sto...

Страница 6: ...oron Scout SN 00256 Wait 120 Minutes for first data The actual status and set up information see below may be displayed by pressing the button again If the first interval has been completed five diffe...

Страница 7: ...6 19 12 34 ALM 10000Bq m INTVL BUZZ OFF TSC 00256 To finish a measurement series keep the Toggle button down and wait for at least four beeps from the buzzer If the button is released the sample will...

Страница 8: ...should be cleared after successful data transmission and storage on hard disk This will save time during the next transmission and avoids redundant data storage Each data record is stored after compl...

Страница 9: ...e accuracy of the measurement From the opposite point of view one could ask How many counts I have to detect to achieve a predefined uncertainty Two items will affect the number of counted decays The...

Страница 10: ...her than the lower limit of the error band of reading 1 Because the true value could be placed within 1350 Bq m and 1469 Bq m the variation of both readings is not statistical significant Example 2 Re...

Страница 11: ...han 16 and therefore the statistical fluctuations have to be derived by the Poisson distribution The stated confidence interval gives the probability that the detected number of decays within the inte...

Страница 12: ...erval 1 255 Minutes adjustable Memory 2047 data records Internal volume chamber approx 60 ml Power supply Battery operation 30 days AC DC adapter 4 5V 0 5A PC Interface Serial USB or RS232 9600 Baud 8...

Отзывы: