Thoron-Scout 

 

 
 

Manual_ThoronScout_EN_02-07-15.doc 

1

User Manual 

Thoron-Scout 

Version 06/2015

 

 

SARAD GmbH  

Tel.: ++49 (0)351 / 6580712 

Wiesbadener Straße 10  

FAX: ++49 (0)351 / 6580718 

D-01159 Dresden 

e-mail: 

[email protected]

 

GERMANY 

Internet: www.sarad.de 

 
 
 
 
 
 
 
 
 

 

 
 

Содержание Thoron-Scout

Страница 1: ...Manual_ThoronScout_EN_02 07 15 doc 1 User Manual Thoron Scout Version 06 2015 SARAD GmbH Tel 49 0 351 6580712 Wiesbadener Stra e 10 FAX 49 0 351 6580718 D 01159 Dresden e mail support sarad de GERMAN...

Страница 2: ...5 CARRYING OUT A MEASUREMENT 6 Adjustment of the sampling interval 7 OPERATION MODES 7 Pump 8 Alarm 8 Fast Slow Mode 8 Sniffing 8 DATA HANDLING 8 Data Storage 8 Data transfer RS232 and USB interface 8...

Страница 3: ...Minutes This time span defines the minimum achievable response time to a Radon concentration step Now the decay chain is continued by the both beta emitters Pb 214 and Bi 214 followed by another alph...

Страница 4: ...tes dislocation during the measurement A display with backlight informs about the actual readings Neither mechanical parts like membrane pumps nor an external power supply are required Therefore use a...

Страница 5: ...is placed at the rear panel of the Thoron Scout Changing the batteries will force you to adjust the re set internal real time clock A concerning message appears on the display of the Thoron Scout Sto...

Страница 6: ...oron Scout SN 00256 Wait 120 Minutes for first data The actual status and set up information see below may be displayed by pressing the button again If the first interval has been completed five diffe...

Страница 7: ...6 19 12 34 ALM 10000Bq m INTVL BUZZ OFF TSC 00256 To finish a measurement series keep the Toggle button down and wait for at least four beeps from the buzzer If the button is released the sample will...

Страница 8: ...should be cleared after successful data transmission and storage on hard disk This will save time during the next transmission and avoids redundant data storage Each data record is stored after compl...

Страница 9: ...e accuracy of the measurement From the opposite point of view one could ask How many counts I have to detect to achieve a predefined uncertainty Two items will affect the number of counted decays The...

Страница 10: ...her than the lower limit of the error band of reading 1 Because the true value could be placed within 1350 Bq m and 1469 Bq m the variation of both readings is not statistical significant Example 2 Re...

Страница 11: ...han 16 and therefore the statistical fluctuations have to be derived by the Poisson distribution The stated confidence interval gives the probability that the detected number of decays within the inte...

Страница 12: ...erval 1 255 Minutes adjustable Memory 2047 data records Internal volume chamber approx 60 ml Power supply Battery operation 30 days AC DC adapter 4 5V 0 5A PC Interface Serial USB or RS232 9600 Baud 8...