User Manual & Datasheet V 2.1
8
Electronic Insert (ICA-3000A)
Please refer to Table 3 for Electronic Insert Specifications.
PARAMETER
VALUE
Working Frequency
500 KHz (Approx)
Tolerable Ambient Temp.
−
20
◦
C to
+80
◦
C
Supply Voltage
15 V DC from Evaluation Unit
Output Signal
3 - 12 V DC
Weight
125 Gms. (Approx.)
Table 3: Electronic Insert
9
Probe
Please refer to Table 4 for Probe Specifications.
PARAMETER
VALUE
Housing
•
RB : Cast Aluminium weather proof powder coated
paint IP-65
•
SQP : Cast Aluminium weather proof powder coated
paint IP-65 (Square Probe Mounted)
•
FP2A : Cast Aluminium weather & flame proof
powder coated paint suitable for gas group IIA & IIB
as per IS 2148
•
SNSR : Pressure Die Cast Aluminium weather proof
IP-66
Cable Entry
1/2"/ 3/4" BSP / NPT, Brass
Cable Gland Type
Single / Double Compression Gland
Mounting
•
Screwed - 1" / 1 1/2" BSP / NPT
•
Flanged - As per your requirement
Insulation
Part / Full PTFE / Ceramic
Working Temp.
Up to
1000
◦
C (Maximum)
Table 4: Probe
Sapcon Instruments Pvt.Ltd.
R
7