-93-
Electrical Adjustment
Grp / No. Item
Function
Range
Initial
Note
7 Flow Ideal Data 8
Flow Sensor Theory Value No.8 Corresponding to Fan Volt No.8
0 - 1023
1023
242
Flow Calib. Fan Volt
(Not Memorized)
0 Flow Calib Fan Volt 1
Flow Sensor Calibration Fan Volt No.1 [V x10]
0 - 255
30*
1 Flow Calib Fan Volt 2
Flow Sensor Calibration Fan Volt No.2 [V x10]
0 - 255
46
2 Flow Calib Fan Volt 3
Flow Sensor Calibration Fan Volt No.3 [V x10]
0 - 255
61
3 Flow Calib Fan Volt 4
Flow Sensor Calibration Fan Volt No.4 [V x10]
0 - 255
84
4 Flow Calib Fan Volt 5
Flow Sensor Calibration Fan Volt No.5 [V x10]
0 - 255
100*
5 Flow Calib Fan Volt 6
Flow Sensor Calibration Fan Volt No.6 [V x10]
0 - 255
255
6 Flow Calib Fan Volt 7
Flow Sensor Calibration Fan Volt No.7 [V x10]
0 - 255
255
7 Flow Calib Fan Volt 8
Flow Sensor Calibration Fan Volt No.8 [V x10]
0 - 255
255
243
Flow Calib Effective Difference
(Not Memorized)
0 Flow Calib Effective Diff 1
Flow Sensor Calibration Range Diff No.1 [ADC]
0 - 1023
160
1 Flow Calib Effective Diff 2
Flow Sensor Calibration Range Diff No.2 [ADC]
0 - 1023
160
2 Flow Calib Effective Diff 3
Flow Sensor Calibration Range Diff No.3 [ADC]
0 - 1023
175
3 Flow Calib Effective Diff 4
Flow Sensor Calibration Range Diff No.4 [ADC]
0 - 1023
185
4 Flow Calib Effective Diff 5
Flow Sensor Calibration Range Diff No.5 [ADC]
0 - 1023
190
5 Flow Calib Effective Diff 6
Flow Sensor Calibration Range Diff No.6 [ADC]
0 - 1023
1023
6 Flow Calib Effective Diff 7
Flow Sensor Calibration Range Diff No.7 [ADC]
0 - 1023
1023
7 Flow Calib Effective Diff 8
Flow Sensor Calibration Range Diff No.8 [ADC]
0 - 1023
1023
244
Flow Clog Threshold
(Not Memorized )
0 Flow Clog Threshold 1
Flow Sensor Clogged Threshold No.1[ADC]
0 - 1023
200
1 Flow Clog Threshold 2
Flow Sensor Clogged Threshold No.2[ADC]
0 - 1023
220
2 Flow Clog Threshold 3
Flow Sensor Clogged Threshold No.3[ADC]
0 - 1023
250
3 Flow Clog Threshold 4
Flow Sensor Clogged Threshold No.4[ADC]
0 - 1023
280
4 Flow Clog Threshold 5
Flow Sensor Clogged Threshold No.5[ADC]
0 - 1023
300
5 Flow Clog Threshold 6
Flow Sensor Clogged Threshold No.6[ADC]
0 - 1023
1023
6 Flow Clog Threshold 7
Flow Sensor Clogged Threshold No.7[ADC]
0 - 1023
1023
7 Flow Clog Threshold 8
Flow Sensor Clogged Threshold No.8[ADC]
0 - 1023
1023
10 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
300
11 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
320
12 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
360
13 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
400
14 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
400
15 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
1023
16 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
1023
17 Flow Inlet Err Threshold 1
Flow Sensor Intake Err Threshold No.1 [ADC]
0 - 1023
1023
245
Filter Roll
0 Scroll One Time
Execute Filter Cleaning (1: Start cleaning once)
0 - 1
0
246
Flow Clog Monitor
(Read only)
0 Real Flow Data
Real Flow Data [ADC] (Read only)
0 - 1023
-
1 Ideal Flow Data
Theory Flow Data (Read only)
0 - 1023
-
2 Flow Difference Data
Flow Difference Data [ADC] (Read only)
-1023 - 1023
-
3 Clog Detect Flow Difference Data Difference Data to judge the Filter Clogged [ADC] (Read only)
0 - 1023
-
4 Warn Detect Flow Difference Data Difference Data to judge the Filter Clogged Warning [ADC] (Read
only)
0 - 1023
-
5 Press Add Data
Add Data by Air Pressure (Read only)
0 - 1023
-
6 Filter Scroll Timer
Timer for Filter Scroll [hour] (Read only)
0 - 32767
-
7 Air Inlet Err Difference Data
Difference to judge the Intake Error
0 - 1023
-
249
Pressure Sensor
0 Pressure Monitor
Pressure Sensor Value [ADC] (Read only)
-
-
1 Pressure Monitor
Pressure Sensor Value with Compensate [ADC] (Read only)
-
-
2 Presss Offset
Offset Value
- 1023 - 1023
0
3 Press calib Effective Diff
Effective Range [ADC]
0 -123
1023
4 Press Sensor Calibration Log
0: Executed & Normal Range (Read ony)
10: Out of Range Minimum
20: Out of Range Maximum
99: Not execute Calibration
-
99
250
Fan Voltage Adjustment
0 Fan 1 Min Adjust (DAC)
DAC Output for Fan Voltage adj.
Adjust Tolerance between DAC and Fan Volt.
*Lamp Mode: Eco
0 - 255
106
1 Fan 1 Max Adjust (DAC)
0 - 255
211
252
Fan Option
1 Safety Switch
Safety Application use
0 - 8
0
Содержание QuaDrive PLC-HF15000L
Страница 189: ... 189 IC Block Diagrams CXD3550 Gamma IC401 CXA7010 S H IC501 IC1501 IC2501 IC1501 IC1551 IC2551 ...
Страница 190: ... 190 IC Block Diagrams FA5501 PFC IC1601 IC1651 HIN202 RS232C Driver IC3801 ...
Страница 192: ... 192 IC Block Diagrams M62393 DAC IC7802 IC5691 IC5692 PIC17F67 Network IC8301 ...
Страница 193: ... 193 IC Block Diagrams TE7783 I O Expander IC1801 ...
Страница 201: ... 201 Parts Location Diagrams KL8 HF15000L00 M01 3 M01 4 Lens shift assembly M01 1 M01 2 M01 2 ...
Страница 206: ... 206 Parts Location Diagrams KL8 HF15000L00 Optical filter LC CS L18 Integrator assembly S06 L08 S06 ...
Страница 207: ... 207 Parts Location Diagrams KL8 HF15000L00 MIrror R assembly S06 L15 Integrator PBS assembly S06 L08 S06 L19 S06 S06 ...
Страница 208: ... 208 Parts Location Diagrams KL8 HF15000L00 Relay lens OUT assembly L06 S06 S06 ...
Страница 209: ... 209 Parts Location Diagrams KL8 HF15000L00 L09 L07 In the Optical unit L14 L14 L05 L21 L13 L11 L20 L10 L12 ...
Страница 210: ... 210 Parts Location Diagrams KL8 HF15000L00 Optical filters assembly Mirror assembly L15 L15 L17 L16 ...
Страница 278: ... KL8A Mar 2011 DC 50 Printed in Japan SANYO Electric Co Ltd ...
Страница 298: ...A18 SCH_KL8A 1 3 4 5 6 7 8 9 10 11 1 13 14 15 16 A B C D E F G H I J K L A B C D E F G H I J K L NO DATA ...