background image

Содержание MPR-311

Страница 1: ...INSTRUCTION MANUAL Pharmaceutical Refrigerator MPR 311 II MPR 3110 H MPR 314DR II MPR 311 DR H ...

Страница 2: ...h the expressed written permission of SANYO 2 The contents of this manual are subject to change without notice 3 Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies SANYO Electric Co Ltd All rights reserved Printed in Japan ...

Страница 3: ...ems and procedures are described so that you can use this unit correctly and safely If the precautions advised are followed this will prevent possible damage to the user and any other person Precautions are illustrated in the following way WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION ...

Страница 4: ...pply outlet with ground earth to prevent electric shock If the power supply outlet is not grounded it will be necessary to install a ground by qualified engineers 2 o o Never ground the unit through a gas pipe water main telephone line or lightening rod Such grounding may cause electric shock in the case of an incomplete circuit Do not insert metal objects such as a pin or a wire into any vent gap...

Страница 5: ...er Do not install the unit where objects may fall on it The electric parts are located in the upper part of this unit Connect the unit to a power source as indicated on the rating label attached to the unit Use of any other voltage or frequency other than that on the rating label may cause fire or electric shock Fix the shelves securely Incomplete installation may cause injury or damage 3 o When r...

Страница 6: ...r touch the condenser directly when the filter is removed for cleaning Surfaces may be hot 4 o o c Always disconnect the power supply plug before moving the unit during transit Take care not to damage the power cord The damaged cord may cause electric shock or fire Dispose of water in the evaporation try completely Spilled water or splashed water may cause current leakage or fire Be careful not to...

Страница 7: ... front glass or frame surface wipe it off with a soft dry cloth l Do not clean the unit with scrubbing brushes acid thinner solvents powdered soap cleanser or hot water These agents can scratch the paint or cause it to peel Plastic and rubber parts can be easily damaged by these materials especially solvents When a neutral detergent is used to clean the unit rinse thoroughly with a cloth soaked in...

Страница 8: ... allows cables to be passed into the cabinet Lock Circuit breaker install position r c L I 6 i i lJ 1 I Caster Drawer _ _ __ L _ _ __ MPR 311DR H For moving of the unit At the time of installation use the height adjusting screw to level the unit Height adjusting screw Adjust the height by screw bolt used to level and install the unit Door switch MPR 311 D H only Temperature recorder install positi...

Страница 9: ... alarm condition Door check lamp DOOR This lamp is lit when the door is open MPR 311 D H MPR 311 DR H Digital temperature display This indicator shows the chamber temperature or set temperature MPR 311D H MPR 311DR H 7 Set button PUSH SET The set temperature is displayed while this button is pressed MPR 311D H MPR 311DR H r v Temperature control knob This knob is for setting a desired temperature ...

Страница 10: ...level and sturdy floor for installation This precaution prevents the unit from tipping Improper installation may result in water spillage or injury from the unit tipping over Install the unit on a sturdy floor to avoid vibration and noise Placing the unit on an unsteady floor will cause vibration and noise 8 A location not prone to high humidity WARNING Do not use the unit outdoors Current Y leaka...

Страница 11: ...ing them counterclockwise to contact them to the floor lb JJ I 0 l e Height adjusting screw Fix the unit Two hooks are attached to the rear of the cabinet Fix the cabinet to the wall with these hooks and rope or chain Hook 9 Grounding WARNING Use an outlet with ground to prevent velectric shock In the case of no outlet with ground it is necessary to provide ground by qualified engineers Never grou...

Страница 12: ...the harness provided with the recorder properly as shown in the figure Remove the cover from the mounting panel Fix the recorder to the mounting panel with two screws 10 Install the mounting panel r J Route the recorder sensor into the chamber by passing through the monitoring hole on the back side After installation power on the unit Check that the temperature alarm is operated properly when the ...

Страница 13: ...unit runs with low temperature less than 5 C Note Never store the material of corrosive such as acid or alkaline of which container is not sealed up completely It may cause poor cool by refrigerant leakage resulted from corrosion of the evaporator I_D_ef_r_os_t_c y_c_le_s______________________________ 1 There are two types of defrost cycles present in this blood bank refrigerator and both cycles o...

Страница 14: ...o 1GOC The chamber temperature is gradually decreased Note An accurate chamber temperature is necessary put a standard thermometer at the center of the chamber Refer to the temperature display of the thermometer for the setting of chamber temperature Table 2 For MPR 311 D H MPR 311 DR H Operation Description 1 Connect the power supply plug to a dedicated power Current chamber temperature is displa...

Страница 15: ...re display is Pulse tone more than 15 C fiashed The chamber temp is High temp alarm lamp is Pulse tone more than about 20 C lit Low temp alarm The chamber temp is less All digits on the temp Pulse tone than DOC display are flashed The chamber temp is less Defrost lamp is lit Pulse tone than about _2 C Door alarm When the door is opened Door check lamp is lit Pulse tone after delay of 30 seconds Th...

Страница 16: ...ectrical insulation and may cause electric shock or short circuit The compressor and other mechanical parts are completely sealed This unit requires absolutely no lubrication There is a fan behind the compressor so be very careful if you stick your hand into this part of the unit Replacement of fluorescent light G Disconnect the power supply plug Remove all shelves or drawers Remove the two screws...

Страница 17: ... temperature too high 6 Is the unit in direct sunlight 7 8 Is the door securely closed Is the door seal damaged or foreign substance inserted between door gaskets ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under Ihe following conditions 1 Indoor use 2 Altitude up to 2000 m 3 Ambient temperature 5 C to 40 C 4 Maximum relative humidity 80 for temperature up to 31 C decre...

Страница 18: ...ture controller Electronic control system Microprocessor control system Thermometer Alcohol type Digital type Fluorescent light White light 20 W x 1 High temp alarm Buzzer High temp alarm lamp ON Buzzer High temp alarm lamp ON Flash of display Low temp alarm Buzzer Defrost lamp ON Buzzer Defrost lamp ON Flash of display Door alarm Door check lamp ON Buzzer after delay of 30 seconds Accessories Key...

Страница 19: ...on of the unit Unit interior DYes DNo No contamination DYes DNo Decontaminated DYes DNo Contaminated DYes DNo Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on DYes DNo b There is some danger see below DYes DNo Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name Pharmaceutica...

Страница 20: ... i t j j I 7FB6P10124000 SANYO Electric Co Ltd Refrigerating Products Division 1 1 1 Sakata Oizumi Machi Ora Gun Gunma 370 0596 Japan Printed in Japan ...

Отзывы: