OPERATING INSTRUCTIONS
9. Start sterilization.
# Press the H H s f l button. A short beep sounds and "°C"lamp
lights to indicate that sterilization is in process.
The digital indicator begins to register, when the temper-
ature inside the chamber reaches 80°C.
The is
lamp starts flashing, when the
temperature inside the chamber reaches 98°C.
Note
• Sterilization can not be started unless
if it is not lit, try to close the door again.
Door lamp
Don't turn the handle, if
flashing.
HIGH PRESSURE
is lit,
lamp is
• Turning the handle and disconnecting the wire connec-
tor, warning buzzer could sound, and §0 error Indica^
tion could be appeared on the digital display, and
lamp, IMfli lamp, liiraisMKMMia and
lamp could be flashed.
(See page 27 "Trouble Shooting Guide".)
HIGH PRESSURE
10. Starting of sterilization timer
• When the temperature inside the chamber reaches the
sterilization temperature setting, the timer operation seg-
ments flash and the timer begins to operate.
# l f the temperature inside the chamber reaches the steri-
lization temperature setting and then drops 0.5°C(0.9°F)
or more below the setting, the timer operation segments
will light continuously and the timer will stop.
If the temperature inside the chamber reaches the steri-
lization temperature setting again the timer will start ope-
rating again.
^—( Supplementary note )-
To determine the remaining sterilization time when the
sterilization timer is operating, press the temperature set-
ting
button and the timer setting |||J5@j| button
at the same time.
(RESET) ( START J^ Short beep
H I G H PRESSURE
38
I MIN I
HIGH PRESSURE HI
C I
Timer operation segments
| HIGH PRESSURE-H |—
(Assuming a sterilization
temperature setting of 121°C)
I 3
•I L*
I
MIN
I
| HIGH PRESSURE-H
IU
- I
MIN
I -
I HIGH PRESSURE-H |-
(10 minutes remaining)
23