background image

Dimensions

[Unit : mm  (inch)]

Specifications

Model

Temperature range

Maximum cooling 
performance

Exterior dimensions
(W x D x H)

Interior dimensions
(W x D x H)

Net weight

Effective capacity

Shelf

Access port

Compressor

Refrigerant

Alarm

Remote alarm contact

Accessories

MDF-U5386S

–50˚C to –86˚C (1˚C increment)

–86˚C 
(Ambient temp. 30˚C)

890 x 875 x 1990mm
(35.0” x 34.4” x 78.3”)

630 x 600 x 1280mm
(24.8” x 23.6” x 50.4”)

Approx. 305kg (672 lbs.)

483 L (17.1 cu.ft.)

Stainless steel, Adjustable, 
3 shelves, W608 x D533mm, 
Max. load: 50kg (110 lbs.)/shelf

17mm diameter, 3 locations 
(back, bottom left/right corner)

Hermetic type, Output: 1100 W 
(high stage side), 1100 W (low 
stage side)

HFC refrigerant

High/low temperature,  
Power failure, Filter check, 
Battery check

Allowable contact capacity: 
DC 30V, 2A

1 set of keys, 1 scraper

Pull-down characteristics

40

20

0

–20

–40

–60

–80

–100

0

2

4

6

8

10

5

0

12

35

25

20

14

15

10

T

e

mperature (˚C)

40

20

0

–20

–40

–60

–80

–100

0

2

4

6

8

10

12

14

T

e

mperature (˚C)

Time (hour)

Time (hour)

Pull-up characteristics 
during power failure

30

14

8

6

4

2

10

12

0

T

ime (hour)

Ambient temperature (˚C)

CO

2

 backup system operation 

characteristics

Central air temp. in cabinet
Ambient temp. 30˚C, No-load

Ambient temp. and maintenance 
of –70˚C by injection of LCO

2

 (30kg)

ACCESSIBILITY

• A new beak style inner 

door latch tightly closes 
the inner door against the 
freezer frame. It also helps make opening 
and closing the door smoother.

• The condenser filter is situated at the

bottom right side of the front panel to 
make filter removal and cleaning easier.

Performance Data

SAFETY

• Alarm lamp and buzzer offer secure 

warning of power failure or abnormal tem-
perature increase.

• High and low temperature warning 

provides an audible and visual alarm when 
the temperature deviates more than 

±

5˚C 

to 

±

20˚C (adjustable) from the set point.

• Alarm ring back function ensures that 

buzzer will resume operation, should alarm 
conditions continue after it is silenced.

• Microprocessor-controlled filter-clog check 

function protects the refrigeration circuit.

• New rugged, one-handed 

outer door latch has a hole to 
allow a padlock to securely 
protect valuable samples.

• Control panel with 

digital display for easy 
operation.

956 (37.6)

890 (35)

18 (0.7)

48

(1.9)

630 (24.8)

1540 (60.6)

450 (17.7)

1990 (78.3)

605.5 (23.8)

105

(4.1)

1280(50.4)

130 (5.1)

130 (5.1)

867 (34.1)

11

(0.4)

1625 (64)

795 (31.3)

140

(5.5)

EVOLUTIONARY DESIGN

• The newly developed refrigeration system 

and the freezer structure offer a quiet 
experimental environment.

• Achieves 10% higher energy conserva-

tion than previous model. 

(For model rated 220V, 50Hz)

• The newly developed back cover combi-

nedwith new aerodynamically designed 
and placed components in the refrigera-
tion compartment 
provide superior 
air flow, making it 
possible to drasti-
cally reduce the 
stress to the freezer 
and contributing to excellent durability.

• Two independent and insulated inner 

doors prevent cold air leakage.

Back 
cover

Compressors

Front

Rear

Air

Options

MDF-U5386S

Distributed by:

©2008 SANYO Printed in Japan 2008.12 MA

SHR137

Features

SANYO Electric Co., Ltd.

Biomedical Division

http://www.sanyo.co.jp/cmg/biomedical

SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for: 

Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001

RoHS (European Restriction of Hazardous Substances directives) compliant

Inventory rack 

IR-220U

Rack Capacity

IR-220U

For 2-inch box

16

IR-312U

For 3-inch box

16

Circular recorder 

MTR-G85

LCO

2

 backup system

CVK-UB2

SANYO DAQ 
(Data Acquisition) 
system 

MTR-480
MTR-2000

Strip chart recorder

MTR-85H

Recorder mounting 
bracket 

MDF-S3085

(for MTR-85H)

•Appearance and specifications are subject to change without notice.

*Cooling performance is indicated by the temperature reached at 

the center of the freezer (at ambient temperature of 30

°

C with no 

load). In order to use the freezer at a stable temperature for a long 
time, it is recommended that the temperature be set to at least 
5

°

C higher than the indicated lowest temperature. 

In addition, depending on the usage conditions, it may not be 
possible to reach the indicated lowest temperature.

Caution:

SANYO guarantees the product under certain warranty conditions. 
SANYO in no way shall be responsible for any loss of content or 
damage to content.

Отзывы: