Dimensions
[Unit : mm (inch)]
Specifications
Model
Temperature range
Maximum cooling
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelf
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
MDF-U5386S
–50˚C to –86˚C (1˚C increment)
–86˚C
(Ambient temp. 30˚C)
890 x 875 x 1990mm
(35.0” x 34.4” x 78.3”)
630 x 600 x 1280mm
(24.8” x 23.6” x 50.4”)
Approx. 305kg (672 lbs.)
483 L (17.1 cu.ft.)
Stainless steel, Adjustable,
3 shelves, W608 x D533mm,
Max. load: 50kg (110 lbs.)/shelf
17mm diameter, 3 locations
(back, bottom left/right corner)
Hermetic type, Output: 1100 W
(high stage side), 1100 W (low
stage side)
HFC refrigerant
High/low temperature,
Power failure, Filter check,
Battery check
Allowable contact capacity:
DC 30V, 2A
1 set of keys, 1 scraper
Pull-down characteristics
40
20
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
5
0
12
35
25
20
14
15
10
T
e
mperature (˚C)
40
20
0
–20
–40
–60
–80
–100
0
2
4
6
8
10
12
14
T
e
mperature (˚C)
Time (hour)
Time (hour)
Pull-up characteristics
during power failure
30
14
8
6
4
2
10
12
0
T
ime (hour)
Ambient temperature (˚C)
CO
2
backup system operation
characteristics
Central air temp. in cabinet
Ambient temp. 30˚C, No-load
Ambient temp. and maintenance
of –70˚C by injection of LCO
2
(30kg)
ACCESSIBILITY
• A new beak style inner
door latch tightly closes
the inner door against the
freezer frame. It also helps make opening
and closing the door smoother.
• The condenser filter is situated at the
bottom right side of the front panel to
make filter removal and cleaning easier.
Performance Data
SAFETY
• Alarm lamp and buzzer offer secure
warning of power failure or abnormal tem-
perature increase.
• High and low temperature warning
provides an audible and visual alarm when
the temperature deviates more than
±
5˚C
to
±
20˚C (adjustable) from the set point.
• Alarm ring back function ensures that
buzzer will resume operation, should alarm
conditions continue after it is silenced.
• Microprocessor-controlled filter-clog check
function protects the refrigeration circuit.
• New rugged, one-handed
outer door latch has a hole to
allow a padlock to securely
protect valuable samples.
• Control panel with
digital display for easy
operation.
956 (37.6)
890 (35)
18 (0.7)
48
(1.9)
630 (24.8)
1540 (60.6)
450 (17.7)
1990 (78.3)
605.5 (23.8)
105
(4.1)
1280(50.4)
130 (5.1)
130 (5.1)
867 (34.1)
11
(0.4)
1625 (64)
795 (31.3)
140
(5.5)
EVOLUTIONARY DESIGN
• The newly developed refrigeration system
and the freezer structure offer a quiet
experimental environment.
• Achieves 10% higher energy conserva-
tion than previous model.
(For model rated 220V, 50Hz)
• The newly developed back cover combi-
nedwith new aerodynamically designed
and placed components in the refrigera-
tion compartment
provide superior
air flow, making it
possible to drasti-
cally reduce the
stress to the freezer
and contributing to excellent durability.
• Two independent and insulated inner
doors prevent cold air leakage.
Back
cover
Compressors
Front
Rear
Air
Options
MDF-U5386S
Distributed by:
©2008 SANYO Printed in Japan 2008.12 MA
SHR137
Features
SANYO Electric Co., Ltd.
Biomedical Division
http://www.sanyo.co.jp/cmg/biomedical
SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for:
Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001
RoHS (European Restriction of Hazardous Substances directives) compliant
Inventory rack
IR-220U
Rack Capacity
IR-220U
For 2-inch box
16
IR-312U
For 3-inch box
16
Circular recorder
MTR-G85
LCO
2
backup system
CVK-UB2
SANYO DAQ
(Data Acquisition)
system
MTR-480
MTR-2000
Strip chart recorder
MTR-85H
Recorder mounting
bracket
MDF-S3085
(for MTR-85H)
•Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at
the center of the freezer (at ambient temperature of 30
°
C with no
load). In order to use the freezer at a stable temperature for a long
time, it is recommended that the temperature be set to at least
5
°
C higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be
possible to reach the indicated lowest temperature.
Caution:
SANYO guarantees the product under certain warranty conditions.
SANYO in no way shall be responsible for any loss of content or
damage to content.