Item MDF-C2156VAN
MDF-C2156VANC
Status monitor function
When ambient temperature is higher than approx. 35
㷄
or lower
than approx. 0
㷄
, ‘Ambient temp is abnormal ‘ is displayed on LCD.
When power supply voltage is lower than approx.170V, ‘The power
supply is abnormal’ is displayed on LCD.
Data log function
z
Recordable for approx. 4weeks in 12min. intervals.
z
Door delay time (more than 2min.) is recordable.
z
Temperature for cascade sensor, filter sensor and AT sensor
are recordable.
z
Log data is transmittable to PC with interface,
MTR-480.(Option)
Self diagnosis function
In the event of any failure among temp. sensor, filter sensor and
cascade sensor;
z
Error code and message are emitted.
z
ALARM lamp flashed, remote alarm contact turns over and
intermittent buzzer emit.
Key lock function
Key lock ‘0’: Key lock is released.
Key lock ‘1’: Key locks
L stage
When the temperature in cascade sensor is -34
㷄
or lower,
compressor (L) turns on.
When the temperature in cascade sensor is -12
㷄
or higher,
compressor (L) turns off.
Overload relay and High pressure switch (L) are controlled.
Compressor
protection
H stage
When the temperature in filter sensor is +60
㷄
or higher,
compressor (H) turns off.
When the temperature in filter sensor is 10
㷄
lower than ambient
temperature, compressor (H) turns on.
Back-up systems
z
Automatic liquid N
2
injection device
z
Temp. setting range;
z
Indication: BACK-UP (Green)
z
Back-up test switch; 8R2021
z
Power switch; Locker switch
z
Back-up systems can be available for 48 hours by full charged
lead storage battery.
㩷
㩷
㩷
㩷
㩷
㩷
Содержание MDF-C2156VAN
Страница 9: ...6 Dimensions ...
Страница 11: ...8 Refrigeration circuits ...
Страница 12: ...9 Refrigeration circuit welding points 䌅䌁㩷 䌅䌂㩷 䌅䌃㩷 䌅䌄㩷 䌅䌆㩷 䌅䌇㩷 Pipe layout ...
Страница 14: ...11 Components on PCB ...
Страница 19: ...16 Wiring Diagram MDF C2156VAN ...
Страница 20: ...17 MDF C2156VANC ...
Страница 21: ...18 Circuit Diagram ...
Страница 22: ...19 CR Circuit Diagram ...