Measure the voltage: Between test points TP‑1,
TP‑2 ,TP‑3 and ground or betweenTP‑4.
(See Figure 28 on page 33)
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
RESULT
RESULT
RESULT
RESULT
RESULT
Test point
TP
Voltage
TP‑1
DC + 5V
TP‑2
DC +12V
TP‑3
DC +12V
TP‑4
DC +12V
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
When removing the cabinet, you must disconnect
the power supply cord from the wall outlet for
your safety. Only the checkout procedure below
needs the power supply on. TAKE GREAT CARE to
avoid possible electrical shock.
For your safety, proceed with the test only
For your safety, proceed with the test only
For your safety, proceed with the test only
For your safety, proceed with the test only
For your safety, proceed with the test only
after removing the wire leads from the primary
after removing the wire leads from the primary
after removing the wire leads from the primary
after removing the wire leads from the primary
after removing the wire leads from the primary
winding of the high voltage transformer.
winding of the high voltage transformer.
winding of the high voltage transformer.
winding of the high voltage transformer.
winding of the high voltage transformer.
CHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURE FOR SWITCHES
CHECKOUT PROCEDURE FOR SWITCHES
Disconnect the lead wires from the switches and check for the continuity of the switches,
connecting an ohm‑meter to its terminals.
CAUTION: After checking the switches, make sure that the interlock monitor
switch is properly connected according to the "CIRCUIT DIAGRAM"
on page 4.
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
CHECKOUT PROCEDURE
Terminals COM and NO
DOOR CLOSE
DOOR CLOSE
DOOR CLOSE
DOOR CLOSE
DOOR CLOSE
SWITCHES
SWITCHES
SWITCHES
SWITCHES
SWITCHES
(See Figure 1 on page 1)
(See Figure 1 on page 1)
(See Figure 1 on page 1)
(See Figure 1 on page 1)
(See Figure 1 on page 1)
DOOR OPEN
DOOR OPEN
DOOR OPEN
DOOR OPEN
DOOR OPEN
INTERLOCK
INTERLOCK
INTERLOCK
INTERLOCK
INTERLOCK
DOOR SENSING
DOOR SENSING
DOOR SENSING
DOOR SENSING
DOOR SENSING
I
II
IINTERLOCK MONITOR
NTERLOCK MONITOR
NTERLOCK MONITOR
NTERLOCK MONITOR
NTERLOCK MONITOR
0
0
0
0
Terminals COM and NO
Primary
winding