— 18 —
Function
Max. Rated Input
Range
DC/AC
110 μA
110.00 μA, 1100.0 μA
DC/AC
110 mA
11.000 mA, 110.00 mA
DC/AC
11 A
11.000 A
1. Never apply a voltage to the measuring terminals.
2. Never apply an input exceeding the maximum rated current.
3. Be sure to connect the meter in series via a load.
WARNING
5-7 Current Measurement (
/
/
)
DCA and ACA
Selection
Remarks:
s
In current measurement, the internal
resistance of the current range is placed
in series and the current drops by this
resistance. Accordingly, its influence
becomes larger in low-resistance
circuits.
s
The AC accuracy guarantee frequency
range is from 45 Hz to 1 kHz.
Be sure that the built-in fuse has not blown.
Power supply
Load
Power supply
Load
Correct measuring method
Wrong measuring method
CAUTION