①
Use the clamp sensor for measurement.
②
Each press of the
button switches the function in
TFRVFODFPG
ACA
→
DCA
→
(DCA+ACA)
→
ACA.
5-6
Current (A) measurements (ACA/DCA/DCA+ACA)
5P JNQSPWF UIF NFBTVSFNFOU BDDVSBDZ QPTJUJPO UIF NFBTVSFE
conductor as close as possible to the center of the clamp sensor.
2. Always clamp around a single wire at a time. Correct current
NFBTVSFNFOUJTJNQPTTJCMFJGNVMUJQMFXJSFTBDBCMFXJUINVMUJQMF
conductors or parallel cords are clamped.
3. Correct measurement may be impossible if the clamp sensor is
not closed completely.
4. Malfunction may occur if the instrument is used in a strong
electromagnetic field.
5. When the 0 adjustment is performed by pressing the
CVUUPO EVSJOH UIF %$ DVSSFOU %$" GVODUJPO NFBTVSFNFOU UIF
displayed value may vary by a few counts due to the reason
EFTDSJCFE BCPWF CVU UIJT JT OPU NBMGVODUJPO 5IF BEKVTUNFOU
should be performed after the orientation of the instrument has
been changed (because of the effects of external magnetic field or
terrestrial magnetism) or when the temperature change is large.
CAUTION