Instruction Manual for R Max
1-888-726-8869
www.sanuvox.com
C. Step by Step LCD Screen Circuit Board Bypass
(Ballast or Circuit Board Validation)
1. Open the R Max ballast box : see step by step instructions of the Ballast Box Opening on page 14 (A).
2. Locate the power input to the circuit board (N,L) as well as the ballast
input (B-N,B-L).
3. Disconnect L and B-L and join them together using twist-on wire
connectors
4. Disconnect N and B-N and join them together using twist-on wire
connectors
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
5. Place the unit on the ground and make sure that you don’t have direct
line of sight with the lamp. Plug the power cord directly in a power
outlet and the UV lamp will light up if both UV lamp and ballast are still
functional.
6. After the test is completed, unplug the unit and replace the wires in their
original position. Black wire from the power cord connects into L and
white wire from the cord connects into N. Black wire going to the ballast
connects into B-L and white wire going to the ballast connects into B-N.