the jumpers are not on the top board, so it may be easier for this procedure to unplug the fan
wiring harness from the digital board to get the wires out of the way. Then, remove the five long
Phillips screws from the digital board as indicated by the arrows in the photo above. Note also
the location of the rectangular box outline in the same photo. This outline indicates the location
of a connector on the bottom of the digital board that attaches to the board below (visible in the
next photo). To remove the digital board you must
carefully pry this connector loose and lift the digital
board away from the camera. Care should be taken
not to pull too strongly on the digital board far from
the connector as this could bend the board and cause
cracks in the delicate traces.
Once the digital board is removed locate the jumper
pins just next to the row of five LEDs that show
through holes in the side of the camera body. The
jumper pins are immediately adjacent to LED02 (see
photo below). There are four pairs of pins. Jumpers
must be placed on the two pairs of pins farthest away
from the LEDs (bottom photo below).
Factory configuration: No jumpers. Jumpers set for Relay Box
Replace the digital board taking care to align all the pins in the connector before applying slight
pressure to seat the pins. Replace the five Phipllips-head screws and plug in the fan wiring
harness. Then replace the back plate.
15
Содержание STXL Advanced Series
Страница 29: ...Appendix E Mechanical Drawings and Backfocus Distances Adapter ring drawings and thread specifications 29...
Страница 30: ...Backfocus Camera without filter wheel 30...
Страница 31: ...Backfocus Camera with Standard Filter Wheel 31...
Страница 32: ...Backfocus Camera with Self Guiding Filter Wheel 32...
Страница 33: ...33...