- 12 -
Enter the boot screen and press the button below the screen
or
set the display value after compensation to 0.
Adjust the contact away from the measuring surface.Click
,
the
system enters the adjustment window, the interface displays G1, records
the display value, adjusts the contact to make the display value change to
about 0 (the value is within
±
30), and locks the contact. At this
point,click
,
the display value becomes 0, the probe compensation
value enters the resetting value, and the adjustment ends.Click
,
enter the measurement interface.
6.3Standard dual-channel sensor adjustment method
(
ZHD
、
ZHF series,except ZHD-1070BC\1090BC\1080BC
)
Enter the boot screen and press the button below the screen
or
set the display value after compensation to 0.
Adjust the contact away from the measuring surface.Click
the
system enters the adjustment window, the interface displays G1, records
the display value, adjusts the lower contact to make the display value
change to about 0 (the value is within
±
30), and locks the contact. At this
point,click
,
the display value becomes 0, the probe compensation
value enters the resetting value, and the lower probe adjustment ends.
Display G2 on the interface, record the size of the display value, adjust the
upper contact to make the display value change to about 0 (within
±
30),
and lock the upper contact. At this point, click
,
the display value
becomes 0, the probe compensation value enters the resetting value, and
the adjustment ends. Click
,
enter the measurement interface.