20
Dimensions
L 140 mm x W 110 mm x H 48 mm
Weight
Approximately 450 g (without bat-
teries)
Cuff size
22 to 32 cm
Permissible operating
conditions
+5°C to +40°C, 30% – 80% relative
air humidity (non-condensing)
Permissible storage
conditions
- 20°C to + 55°C, 10% – 93% relative
air humidity, 500 – 1060 hPa ambient
pressure
Power supply
4 x 1.5 V
AA batteries (alkali-
ne type LR06)
Battery life
For approx. 200 measurements, de-
pending on levels of blood pressure
and pump pressure
Accessories
Instructions for use, 4 x 1.5 V AA
batteries (alkaline type LR06), storage
pouch
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
Technical information is subject to change without notification
to allow for updates.
• This device complies with European Standard EN60601-1-2
and is subject to particular precautions with regard to electro-
magnetic compatibility. Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Service
address or found at the end of the instructions for use.
• This device corresponds to the EU Medical Devices Directive
93/42/EC, the German Medical Devices Act (Medizinproduk-
tgesetz) and the standards EN 1060-1 (non-invasive sphyg-
momanometers, Part 1: General requirements), EN 1060-3
(non-invasive sphygmomanometers, Part 3: Supplemen-
tary requirements for electro-mechanical blood pressure
measuring systems) and IEC 80601-2-30 (Medical electrical
equipment – Part 2 – 30: Particular requirements for the basic
safety and essential performance of automated non-invasive
sphygmomanometers).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life. If using the device for commercial medical purposes, it
must be regularly tested for accuracy by appropriate means.
Precise instructions for checking accuracy may be requested
from the service address.