: errors during measuring,
•
: the pump pressure is higher than 300 mmHg,
•
: errors in the recorded measurement data,
•
: there is a system or unit error.
In such cases, repeat the measurement. Ensure that you do
not move or speak.
If necessary, reinsert or replace the batteries.
8. Cleaning and Care
• Clean your blood pressure monitor carefully using a slightly
damp cloth only.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.
9. Technical details
Model no.
SBC 25
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the wrist
Measurement
range
Cuff pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
Pulse 40 –199 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
3 x 40 memory spaces
Dimensions
L 70 mm x W 72 mm x H 28,6 mm
Weight
Approx. 112 g (without batteries)
Cuff size
135 to 195 mm
Permissible operat-
ing conditions
+ 5 °C to +40 °C, 15 – 93 % relative air
humidity (non-condensing)
Permissible storage
conditions
-25 °C to +70 °C, ≤ 93 % relative air
humidity, 700 –1060 hPa ambient pres-
sure
Power supply
2 x 1,5 V
AAA batteries
Battery life
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA batter-
ies, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notifica-
tion to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to elec-
tromagnetic compatibility (EMC). Please note that port-
able and mobile HF communication systems may inter-
fere with this unit. More details can be requested from the
19