12
8. Transferring measured values
Bluetooth
®
Smart transfer
It is also possible to transfer the measured values saved on
the device to your smartphone using Bluetooth
®
Smart. You
will need the “Sanitas HealthCoach” app for this. The app is
available from the App Store.
Proceed as follows to manually transfer values:
Select the memory mode (chapter 7 “Saving, displaying and
deleting measured values”). As soon as the average value
of all the measured values is displayed, a Bluetooth
®
con
-
nection is established. The Bluetooth
®
symbol is displayed
during transfer.
More information:
www.sanitas-online.de
9. Error messages/troubleshooting
In the event of errors, the error message _ appears on the
screen.
Error messages may appear if:
•
It was not possible to correctly record the pulse:
1
;
•
You speak or move during the measurement:
2
;
•
The cuff is fastened too tightly or loosely:
3
;
•
Errors occur during the measurement:
4
;
•
The pump pressure is higher than 300 mmHg:
5
;
•
The batteries are almost empty:
6
.
•
The data could not be sent via Bluetooth:
7
.
In such cases, repeat the measurement. Ensure that you do
not move or speak.
If necessary, reinsert or replace the batteries.
10. Cleaning and storing the unit
•
Clean your blood pressure monitor carefully using a
slightly damp cloth only.
•
Do not use any cleaning agents or solvents.
•
Under no circumstances should you hold the device
under water, as this can cause liquid to enter and damage
the device.
•
If you store the unit, no heavy objects should be placed
on top of it. Remove the batteries. The cuff line should not
be bent sharply.
11. Technical specifications
Model No.
MD3600
Measurement
method
Oscillometric, non-invasive blood
pressure measurement on the
upper arm
Measurement
range
Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute